background image

mA Calibrator 

User-Replaceable Parts 

Replacing the Battery 

XW

 Warning 

To avoid false readings, which could lead to 
electric shock or injury, replace the battery 
as soon as 

M

 (low battery indicator) 

appears on the display. 

Do not remove or install the battery within 
the Ex-hazardous area. 

Use only a single 9 V alkaline battery, 
properly installed, to power the  
mA Calibrator. See Table 2.  

To replace the battery (refer to Figure 1 and Table 2): 

1.

 

Remove the test probes from the input signal. 

2.

 

Press 

D

 to turn the mA Calibrator OFF. 

3.

 

Remove the test leads from the input terminals. 

4.

 

Remove the red holster. 

5.

 

Lift off the battery door on the back of the mA 

Calibrator as shown. 

6.

 

Remove the battery. 

7.

 

Insert the replacement battery and re-install the 

battery door. Make sure it is securely in place. 

8.

 

Return the mA Calibrator to its red holster. 

 

aqv07f.eps 

Figure 1. Replacing the Battery 

User-Replaceable Parts 

User-replaceable parts and accessories are shown in 

Figure 2 and Table 1. 

1

2

3

6

5

4

7

 

aqv10f.eps 

Figure 2. User-Replaceable Parts 

Table 1. User-Replaceable Parts 

Item

Part Description 

Fluke Part No.

Qty.

A

 Non-skid 

foot 

885884 

B

 

Battery, 9 volt, 
Alkaline 

822270 or 
see Table 2 

C

 Battery 

door 

665106 

D

 

Red holster, with bail 

2040228 

E

 Alligator 

clips 

AC72 

F

 

Test lead set 

TL75 

G

 

Users Manual on CD  

2053979 

Table 2. Approved Batteries 

Battery Description 

Manufacturer Type 

Alkaline, 9 volt 

Duracell 

6LR61 

Alkaline Ultra, 9 volt 

Duracell 

6LR61 

Professional Alkaline 
Battery Procell, 9 volt 

Duracell 6LR61 

Alkaline Energizer, 9 volt 

Eveready 

6LR61 

Alkaline Power Line 
Industrial Battery, 9 volt 

Panasonic 6LR61 

Alkaline, 9 volt 

Daimon 

6LR61 

Содержание 707Ex

Страница 1: ...ete operating instructions refer to the 707Ex Users Manual located on the CD that came with the mA Calibrator Warranty This Fluke product comes with a 3 year warranty for material and workmanship For warranty information refer to the 707Ex Users Manual that came with the mA Calibrator Contacting Fluke To contact Fluke call one of the following telephone numbers Technical Support USA 1 800 44 FLUKE...

Страница 2: ... to 18 C and 28 C to 50 C Relative Humidity 95 up to 30 C 75 up to 40 C 45 up to 50 C Vibration Random 2 g 5 to 500 Hz Shock 1 meter drop test Safety Compliance Complies with IEC 61010 1 95 CAT I 28 V CSA C22 2 No 1010 92 NRTL ANSI ISA S82 02 01 1994 Directive 94 9 EG and p NEC 500 Uo 27 6 V Io 96 13 mA Co 76 nF Lo 2 5 mH Ui 30 V Ii 24 mA Ci 10 nF Li 0 mH CE Complies with EN61010 1 and EN61326 Pow...

Страница 3: ...t is used Measurement transients are defined in IEC1010 1 as 2 µs rise time with a 50 µs duration at 50 of the maximum amplitude height Measurement Category I CAT I is defined for measurements performed on circuits not directly connected to the mains Do not use in a damp or wet environment Faults and Damage XW Warning Applying a voltage greater than 28 V to the input of the mA Calibrator invalidat...

Страница 4: ...0460 2 P Conforms to European Union requirements F Direct current Do not dispose of this product as unsorted municipal waste Go to Fluke s website for recycling information Conforms to relevant Australian standards Maintenance XW Warning To avoid electric shock personal injury or damage to the mA Calibrator Do not service this product To maintain the integrity of the mA Calibrator in explosive atm...

Страница 5: ...or handling them 3 Using the procedures packaging and bench techniques that are recommended The following practices should be followed to minimize damage to S S static sensitive devices 1 MINIMIZE HANDLING 2 KEEP PARTS IN ORIGINAL CONTAINERS UNTIL READY FOR USE 3 DISCHARGE PERSONAL STATIC BEFORE HANDLING DEVICES USE A HIGH RESIS TANCE GROUNDING WRIST STRAP 4 HANDLE S S DEVICES BY THE BODY static a...

Страница 6: ...VE EDGES AND NEVER TOUCH OPEN EDGE CONNECTOR EXCEPT AT STATIC FREE WORK STATION PLACING SHORTING STRIPS ON EDGE CONNECTOR HELPS PROTECT INSTALLED S S DEVICES 9 HANDLE S S DEVICES ONLY AT A STATIC FREE WORK STATION 10 ONLY ANTI STATIC TYPE SOLDER SUCKERS SHOULD BE USED 11 ONLY GROUNDED TIP SOLDERING IRONS SHOULD BE USED PORTIONS REPRINTED WITH PERMISSION FROM TEKTRONIX INC AND GERNER DYNAMICS POMON...

Страница 7: ...ery and re install the battery door Make sure it is securely in place 8 Return the mA Calibrator to its red holster aqv07f eps Figure 1 Replacing the Battery User Replaceable Parts User replaceable parts and accessories are shown in Figure 2 and Table 1 1 2 3 6 5 4 7 aqv10f eps Figure 2 User Replaceable Parts Table 1 User Replaceable Parts Item Part Description Fluke Part No Qty A Non skid foot 88...

Страница 8: ...C Amps DCI 6 Refer to Table 4 to verify the readings on the HP 3458A for the following tests No adjustment is necessary for test 1 Test 2 1 Press B twice to select 12 000 mA 2 Verify the readings Test 3 1 Press B twice to select 20 000 mA 2 Press and rotate the knob until 24 000 mA is displayed 3 Verify the readings DC Current Measurement Mode Tests 4 6 1 Press the A key twice to put the mA Calibr...

Страница 9: ...and set output to STANDBY 7 Turn the mA Calibrator OFF and disconnect the mA Calibrator from the 5520A 8 The performance verification tests are now complete 9 If the mA Calibrator failed any of these tests calibration adjustment or repair is required Calibration Adjustment Perform the following calibration adjustment procedure if the mA Calibrator fails the performance verification test Calibratio...

Страница 10: ...st 6 Disconnect the HP 3458A from the mA Calibrator 7 Connect the test leads from the AUX terminals of the Fluke 5520A to the terminals on the mA Calibrator black to COM and red to 8 The mA Calibrator will display MEASURE and 0 000 mA is displayed On the secondary display the last 3 digits of the A D reading are shown Set the 5520A to output 0 000 mA into the mA Calibrator Press C 9 After a moment...

Отзывы: