Fluke 707Ex 
Calibration Information
 

secondary display. If the knob is not turned 

within 2 seconds, the unit returns to the normal 

operation mode. 

2.

 

The mA Calibrator adjustment procedure is 

password protected. At this point, a password is 

needed to go forward. "

PAS

" is displayed on the 

main display and "

000

" on the secondary display.  

Enter the password "

297

" as follows: 

1.

 

Rotate the knob to add a 2 in the LCD of 

the secondary display push and release 

the knob. 

2.

 

Rotate the knob to get 9 into the LCD, 

push and release.  

3.

 

Rotate the knob to get a 7, push and 

release. 

If the password is correct, you'll be led through all 

the adjust modes (source, mA read, V read), 

otherwise the unit returns to normal operation. In 

the adjust mode, the main display will show 

"

1.000 mA

"

3.

 

Connect the mA Calibrator 

+

 terminal to the HP 

3458A 

input, and the 

COM

 terminal to the HP 

3458A LO input. The mA Calibrator will be in 

mA source cal and "

1.000 mA

" is displayed. On 

the secondary LCD, the last 3 digits from the 

knob value are shown. Adjust the mA output by 

tuning the knob until "

1.000 mA

" shows on the 

HP 3458A. 

4.

 

Press 

C

, "

22.000 mA

" is displayed. Adjust the 

mA output by turning the knob until "

22.000 mA

is shown on the HP 3458A. 

5.

 

Press 

A

 to go to mA measure adjust. 

6.

 

Disconnect the HP 3458A from the mA 

Calibrator. 

7.

 

Connect the test leads from the 

AUX

 

terminals of 

the Fluke 5520A to the terminals on the mA 

Calibrator  (black to 

COM

 

and red to 

+

). 

8.

 

The mA Calibrator will display "

MEASURE

" and 

0.000 mA is displayed. On the secondary display, 

the last 3 digits of the A/D reading are shown. Set 

the 5520A to output 0.000 mA into the mA 

Calibrator. Press 

C

.

 

9.

 

After a moment, the mA Calibrator will display 

24.000 mA. Set the 5520A to output 24.000 mA 

into the mA Calibrator.  

10.

 

Press 

A

 to go to V measure adjust. 

11.

 

Connect the test leads from the output 

NORMAL 

terminals of the Fluke 5520A to the input 

terminals on the mA Calibrator. 

12.

 

The mA Calibrator will display "

MEASURE

" and 

0.000 V. The Secondary display will show the 

last 3 digits of the A/D reading. Set the 5520A to 

output 0.000 V into the mA Calibrator. 

13.

 

Press 

C

. After a moment, 25.000 V will 

display on the mA Calibrator. Set the 5520A to 

output 25.000 V into the mA Calibrator. 

14.

 

Press 

A

 to complete the adjustment. The 

calibration counter is incremented and the 

calibration data is stored. The calibration count 

will be on the main display and "

CAL

" on the 

secondary display for 2 seconds.  

The calibration adjustment procedure is complete. 

To see the value of the calibration counter without 

calibrating the unit, hold down 

B

 

at power up. The 

software revision number is displayed for 2 seconds 

followed by the calibration counter for 2 seconds. 

 

 

 

Содержание 707Ex

Страница 1: ...ete operating instructions refer to the 707Ex Users Manual located on the CD that came with the mA Calibrator Warranty This Fluke product comes with a 3 year warranty for material and workmanship For warranty information refer to the 707Ex Users Manual that came with the mA Calibrator Contacting Fluke To contact Fluke call one of the following telephone numbers Technical Support USA 1 800 44 FLUKE...

Страница 2: ... to 18 C and 28 C to 50 C Relative Humidity 95 up to 30 C 75 up to 40 C 45 up to 50 C Vibration Random 2 g 5 to 500 Hz Shock 1 meter drop test Safety Compliance Complies with IEC 61010 1 95 CAT I 28 V CSA C22 2 No 1010 92 NRTL ANSI ISA S82 02 01 1994 Directive 94 9 EG and p NEC 500 Uo 27 6 V Io 96 13 mA Co 76 nF Lo 2 5 mH Ui 30 V Ii 24 mA Ci 10 nF Li 0 mH CE Complies with EN61010 1 and EN61326 Pow...

Страница 3: ...t is used Measurement transients are defined in IEC1010 1 as 2 µs rise time with a 50 µs duration at 50 of the maximum amplitude height Measurement Category I CAT I is defined for measurements performed on circuits not directly connected to the mains Do not use in a damp or wet environment Faults and Damage XW Warning Applying a voltage greater than 28 V to the input of the mA Calibrator invalidat...

Страница 4: ...0460 2 P Conforms to European Union requirements F Direct current Do not dispose of this product as unsorted municipal waste Go to Fluke s website for recycling information Conforms to relevant Australian standards Maintenance XW Warning To avoid electric shock personal injury or damage to the mA Calibrator Do not service this product To maintain the integrity of the mA Calibrator in explosive atm...

Страница 5: ...or handling them 3 Using the procedures packaging and bench techniques that are recommended The following practices should be followed to minimize damage to S S static sensitive devices 1 MINIMIZE HANDLING 2 KEEP PARTS IN ORIGINAL CONTAINERS UNTIL READY FOR USE 3 DISCHARGE PERSONAL STATIC BEFORE HANDLING DEVICES USE A HIGH RESIS TANCE GROUNDING WRIST STRAP 4 HANDLE S S DEVICES BY THE BODY static a...

Страница 6: ...VE EDGES AND NEVER TOUCH OPEN EDGE CONNECTOR EXCEPT AT STATIC FREE WORK STATION PLACING SHORTING STRIPS ON EDGE CONNECTOR HELPS PROTECT INSTALLED S S DEVICES 9 HANDLE S S DEVICES ONLY AT A STATIC FREE WORK STATION 10 ONLY ANTI STATIC TYPE SOLDER SUCKERS SHOULD BE USED 11 ONLY GROUNDED TIP SOLDERING IRONS SHOULD BE USED PORTIONS REPRINTED WITH PERMISSION FROM TEKTRONIX INC AND GERNER DYNAMICS POMON...

Страница 7: ...ery and re install the battery door Make sure it is securely in place 8 Return the mA Calibrator to its red holster aqv07f eps Figure 1 Replacing the Battery User Replaceable Parts User replaceable parts and accessories are shown in Figure 2 and Table 1 1 2 3 6 5 4 7 aqv10f eps Figure 2 User Replaceable Parts Table 1 User Replaceable Parts Item Part Description Fluke Part No Qty A Non skid foot 88...

Страница 8: ...C Amps DCI 6 Refer to Table 4 to verify the readings on the HP 3458A for the following tests No adjustment is necessary for test 1 Test 2 1 Press B twice to select 12 000 mA 2 Verify the readings Test 3 1 Press B twice to select 20 000 mA 2 Press and rotate the knob until 24 000 mA is displayed 3 Verify the readings DC Current Measurement Mode Tests 4 6 1 Press the A key twice to put the mA Calibr...

Страница 9: ...and set output to STANDBY 7 Turn the mA Calibrator OFF and disconnect the mA Calibrator from the 5520A 8 The performance verification tests are now complete 9 If the mA Calibrator failed any of these tests calibration adjustment or repair is required Calibration Adjustment Perform the following calibration adjustment procedure if the mA Calibrator fails the performance verification test Calibratio...

Страница 10: ...st 6 Disconnect the HP 3458A from the mA Calibrator 7 Connect the test leads from the AUX terminals of the Fluke 5520A to the terminals on the mA Calibrator black to COM and red to 8 The mA Calibrator will display MEASURE and 0 000 mA is displayed On the secondary display the last 3 digits of the A D reading are shown Set the 5520A to output 0 000 mA into the mA Calibrator Press C 9 After a moment...

Отзывы: