6102 Micro-Bath User’s Guide
Operating Parameters
40
16
New Stir Speed setting
Press “SET” to accept the new Stir Speed.
The stir motor speed needs to be varied for best stability. Table 4 on page 28 shows
nominal settings for several fluids.
High Limit
9.10.2
The High Limit Parameter adjusts the upper set-point temperature. The factory default
and maximum temperature are set to 200°C. For safety, a user can ad just the High
Limit down so the maximum temperature set-point is restricted.
HL
High Limit parameter
Press “SET” to enable adjustment of the High Limit parameter.
HL
Flashes “HL” and then displays the setting
H=126
Current HL setting
Adjust the HL parameter using the “UP” or “DOWN” keys.
H=90
New HL setting
To store the new setting, press “SET”. Press “EXIT” to continue without storing the
new value.
Cutout Reset Mode
9.10.3
The Cutout Reset Mode determines whether the cutout resets automatically when the
tank temperature drops to a safe value or must be manually reset by the operator.
CtorSt
Cutout Reset Mode parameter
Press “SET” to enable adjustment of the Cutout Reset Mode parameter. Nor mally, the
cutout is set for manual mode.
CtorSt
Flashes “CtorSt” and then displays the setting
rst
Cutout set for manual reset
To change to automatic reset mode press “UP” and then “SET”.
Auto
Cutout set for automatic reset
1.888.475.5235
Fluke-Direct
.com
Содержание 6102
Страница 16: ...1 888 475 5235 info Fluke Direct com Fluke Direct com ...
Страница 18: ...1 888 475 5235 info Fluke Direct com Fluke Direct com ...
Страница 38: ...1 888 475 5235 info Fluke Direct com Fluke Direct com ...
Страница 58: ...1 888 475 5235 info Fluke Direct com Fluke Direct com ...
Страница 62: ...1 888 475 5235 info Fluke Direct com Fluke Direct com ...
Страница 64: ...1 888 475 5235 info Fluke Direct com Fluke Direct com ...