6102 Micro-Bath User’s Guide
Safety Information
2
Symbol
Description
C-TICK Australian EMC mark
The European Waste Electrical and Electronic Equipment (WEEE) Directive (2002/96/
EC) mark.
OVERVOLTAGE (Installation) CATEGORY II, Pollution Degree 2 per IEC1010-1 re fers to
the level of Impulse Withstand Voltage protection provided. Equipment of OVERVOLTAGE
CATEGORY II is energy-consuming equipment to be supplied from the fixed installation.
Examples include household, office, and laboratory appliances.
Safety Information
1.2
Use this instrument only as specified in this manual. Otherwise, the protection
provided by the instrument may be impaired.
The following definitions apply to the terms “Warning” and “Caution”.
“Warning” identifies conditions and actions that may pose hazards to the user.
L
“Caution” identifies conditions and actions that may damage the instru ment
L
being used.
Warnings
1.2.1
To avoid personal injury, follow these guidelines.
GENERAL
DO NOT
use the instrument for any application other than calibration work. The
instrument was designed for temperature calibration. Any other use of the unit may
cause unknown hazards to the user.
DO NOT
overfill the bath. Overflowing extremely cold or hot fluid may be harmful to
the operator. See Section 5.3, Bath Preparation and Filling, for specific instructions.
DO NOT
install access cover without holes onto a bath that is energized. Dangerous
pressures may result from fluids vaporizing.
Calibration Equipment should only be used by Trained Personnel.
If this equipment is used in a manner not specified by the manufacturer, the protection
provided by the equipment may be impaired.
Before initial use, or after transport, or after storage in humid or semi-humid
environments, or anytime the instrument has not been energized for more than 10
days, the instrument needs to be energized for a “dry-out” period of 2 hours before
it can be assumed to meet all of the safety requirements of the IEC 61010-1. If the
product is wet or has been in a wet environment, take necessary measures to remove
moisture prior to applying power such as storage in a low humidity temperature
chamber operating at 50°C for 4 hours or more.
1.888.475.5235
Fluke-Direct
.com
Содержание 6102
Страница 16: ...1 888 475 5235 info Fluke Direct com Fluke Direct com ...
Страница 18: ...1 888 475 5235 info Fluke Direct com Fluke Direct com ...
Страница 38: ...1 888 475 5235 info Fluke Direct com Fluke Direct com ...
Страница 58: ...1 888 475 5235 info Fluke Direct com Fluke Direct com ...
Страница 62: ...1 888 475 5235 info Fluke Direct com Fluke Direct com ...
Страница 64: ...1 888 475 5235 info Fluke Direct com Fluke Direct com ...