6.4
Thermal EMF
Two factors contribute to thermal EMF, chemical consistency and physical con-
sistency. Variations in chemical structure due to impurities can contribute to
thermal EMF. Also discrepancies in crystal structure can contribute to thermal
EMF. These factors are minimized by annealing the full length of wire before
construction of the PRT.
Likewise, connection to extension lead wires and readout instruments can be a
source of thermal EMF. The thermal EMF is caused by a difference in tempera-
ture between two connections. If the two connections are the same temperature,
there will be little or no thermal EMF effects. However, if there is a substantial
temperature difference between connections, the thermal EMF effects will be
significant. Therefore, cover or insulate any exposed bridge or galvanometer
terminals to lessen the source of error. The effects of thermal EMF can be can-
celed by using an AC bridge or a DC bridge with reversible current.
5624 Platinum Resistance Thermometer
User’s Guide
18