5502E
Getting Started
30
Table 7. Verification Tests for AC Voltage (Normal)
Range
Output
Frequency
Lower Limit
Upper Limit
32.999 mV
3.000 mV
45 Hz
2.977 mV
3.022 mV
32.999 mV
3.000 mV
10 kHz
2.977 mV
3.022 mV
32.999 mV
30.000 mV
9.5 Hz
28.350 mV
31.650 mV,
32.999 mV
30.000 mV
10 Hz
29.944 mV
30.056 mV
32.999 mV
30.000 mV
45 Hz
29.956 mV
30.044 mV
32.999 mV
30.000 mV
1 kHz
29.956 mV
30.044 mV
32.999 mV
30.000 mV
10 kHz
29.956 mV
30.044 mV
32.999 mV
30.000 mV
20 kHz
29.944 mV
30.056 mV
32.999 mV
30.000 mV
50 kHz
29.932 mV
30.068 mV
32.999 mV
30.000 mV
100 kHz
29.877 mV
30.123 mV
32.999 mV
30.000 mV
450 kHz
29.715 mV
30.285 mV
329.999 mV
33.000 mV
45 Hz
32,970 mV
33.029 mV
329.999 mV
33.000 mV
10 kHz
32.970 mV
33.029 mV
329.999 mV
300.000 mV
9.5 Hz
283.350 mV
316.650 mV
329.999 mV
300.000 mV
10 Hz
299.917 mV
300.083 mV
329.999 mV
300.000 mV
45 Hz
299.893 mV
300.107 mV
329.999 mV
300.000 mV
1 kHz
299.983 mV
300.107 mV
329.999 mV
300.000 mV
10 kHz
299.983 mV
300.107 mV
329.999 mV
300.000 mV
20 kHz
299.782 mV
300.218 mV
329.999 mV
300.000 mV
50 kHz
299.702 mV
300.298 mV
329.999 mV
300.000 mV
100 kHz
299.311 mV
300.689 mV
329.999 mV
300.000 mV
500 kHz
298.470 mV
301.530 mV
3.29999 V
0.33000 V
45 Hz
0.32984 V
0.33015 V
3.29999 V
0.33000 V
10 kHz
0.32984 V
0.33015V
3.29999 V
3.00000 V
9.5 Hz
2.83500 V
3.16500 V
3.29999 V
3.00000 V
10 Hz
2.99868 V
3.00132 V
3.29999 V
3.00000 V
45 Hz
2.99910 V
3.00090 V
3.29999 V
3.00000 V
1 kHz
2.99910V
3.00090 V
3.29999 V
3.00000 V
10 kHz
2.99910 V
3.00090 V
3.29999 V
3.00000 V
20 kHz
2.99817 V
3.00183 V
3.29999 V
3.00000 V
50 kHz
2.99745 V
3.00255 V
Содержание 5502E
Страница 4: ......
Страница 8: ...5502E Getting Started iv...
Страница 10: ...5502E Getting Started vi...
Страница 48: ...5502E Getting Started 38...