Multi-Product Calibrator
Additional Specifications
21
Temperature Calibration (RTD)
RTD Type
Range
°
C
[1]
Absolute Uncertainty tcal
±
5
°
C
±
°
C
[2]
RTD Type
Range
°
C
[1]
Absolute Uncertainty
tcal
±
5
°
C
±
°
C
[2]
90 Day
1 Year
90 Day
1 Year
Pt 385,
100
Ω
-200 to -80
0.04
0.05
Pt 385,
500
Ω
-200 to -80
0.03
0.04
-80 to 0
0.05
0.05
-80 to 0
0.04
0.05
0 to 100
0.07
0.07
0 to 100
0.05
0.05
100 to 300
0.08
0.09
100 to 260
0.06
0.06
300 to 400
0.09
0.10
260 to 300
0.07
0.08
400 to 630
0.10
0.12
300 to 400
0.07
0.08
630 to 800
0.21
0.23
400 to 600
0.08
0.09
Pt 3926,
100
Ω
-200 to -80
0.04
0.05
600 to 630
0.09
0.11
-80 to 0
0.05
0.05
Pt 385,
1000
Ω
-200 to -80
0.03
0.03
0 to 100
0.07
0.07
-80 to 0
0.03
0.03
100 to 300
0.08
0.09
0 to 100
0.03
0.04
300 to 400
0.09
0.10
100 to 260
0.04
0.05
400 to 630
0.10
0.12
260 to 300
0.05
0.06
Pt 3916,
100
Ω
-200 to -190
0.25
0.25
300 to 400
0.05
0.07
-190 to -80
0.04
0.04
400 to 600
0.06
0.07
-80 to 0
0.05
0.05
600 to 630
0.22
0.23
0 to 100
0.06
0.06
PtNi 385,
120
Ω
(Ni120)
-80 to 0
0.06
0.08
100 to 260
0.06
0.07
0 to 100
0.07
0.08
260 to 300
0.07
0.08
100 to 260
0.13
0.14
300 to 400
0.08
0.09
Cu 427
10
Ω
[3]
-100 to 260
0.3
0.3
400 to 600
0.08
0.10
600 to 630
0.21
0.23
Pt 385,
200
Ω
-200 to -80
0.03
0.04
-80 to 0
0.03
0.04
0 to 100
0.04
0.04
100 to 260
0.04
0.05
260 to 300
0.11
0.12
300 to 400
0.12
0.13
400 to 600
0.12
0.14
600 to 630
0.14
0.16
[1]
Resolution is 0.003
°
C
[2]
Applies for COMP OFF (to the 5502E Calibrator front panel NORMAL terminals) and 2-wire and 4-wire compensation.
[3]
Based on MINCO Application Aid No. 18
Additional Specifications
The subsequent paragraphs provide additional specifications for the 5502E Calibrator ac voltage and ac current functions.
These specifications are valid after allowing a warm-up period of 30 minutes, or twice the time the 5502E has been turned
off. All extended range specifications are based on performing the internal zero-cal function at weekly intervals, or when
the ambient temperature changes by more than 5
°
C.
Frequency
Frequency Range
Resolution
1-Year Absolute Uncertainty, tcal
±
5
°
C
±
(ppm + mHz)
Jitter
0.01 to 119.99 Hz
0.01 Hz
25 + 1
2
μ
s
120.0 to 1199.9 Hz
0.1 Hz
25 + 1
2
μ
s
1.2 to 11.999 kHz
1 Hz
25 + 1
2
μ
s
12 to 119.99 kHz
10 Hz
25 + 15
140 ns
120.0 to 1199.9 kHz
100 Hz
25 + 15
140 ns
1.2 to 2.000 MHz
1 kHz
25 + 15
140 ns
Содержание 5502E
Страница 4: ......
Страница 8: ...5502E Getting Started iv...
Страница 10: ...5502E Getting Started vi...
Страница 48: ...5502E Getting Started 38...