Table 4-2. Performance Tests (Voltage, Resistance, and Frequency) (cont)
FUNCTION
RANGE
INPUT
FREQUENCY
DISPLAY ACCURACY
(1 Year, 18-28
°
C)
LEVEL
MIN
MAX
300
Ω
3 k
Ω
3 k
Ω
30 k
Ω
300 k
Ω
3 M
Ω
10 M
Ω
short
100
Ω
short
1 k
Ω
10 k
Ω
100 k
Ω
1 M
Ω
10
0.00
99.92
0.0000
0.9992
9.992
99.92
0.9992
9.986
0.09
100.15
0.0003
1.0009
10.008
100.08
1.0008
10.014
* Optional test point if standards available.
Note
All channels (0 through 20) can accommodate 2-terminal resistance
measurements. Channel 0, with only two connections, cannot be used for 4-
terminal measurements. Four-terminal resistance measurements can be
defined for channels 1 through 10 only. Channels 11 through 20 are used,
as required, for 4-terminal to provide the additional two connections. For
example, a 4-terminal set up on channel 1 uses channels 1 and 11, each
channel providing two connections.
Frequency
90 kHz
10 kHz/2V p-p
9.994
10.006
8.
Connect a cable from the Output VA HI and LO of the 5700A to the Input Module
test leads (observe proper polarity).
9.
Select the VDC function and 300-volt range on Hydra and apply 0V dc from the
5700A. Then apply 290V dc input from the 5700A. Ensure the display reads
between the minimum and maximum values as shown in Table 4-2 for the 0 and
290V dc input levels.
Note
Channels 0, 1, and 11 can accommodate a maximum input of 300V dc or
ac. However, the maximum input for all other channels can only be 150V
dc or ac.
10.
With the exception of the selected voltage range and input voltagefrom the 5700A,
repeat steps 1 through 9 for each remaining InputModule channel (2 through 20).
Channels 2 through 10 and 12 through20 should be configured for the 150V dc range
and an input voltageof 150 volts.
4-6.
Thermocouple Measurement Range Accuracy Test
Verify that the Accuracy Verification Test for channel 0 meets minimum acceptable
levels before performing this test.
Thermocouple temperature measurements are accomplished using the Hydra internal
100 mV and 1V dc ranges. (The ranges are not configurable by the operator.) This
procedure provides the means to test these ranges.
Testing the 100 mV and 1V dc ranges requires computer interfacing with a host
(terminal or computer). The host must send commands to select these ranges. These
ranges cannot be selected from the Hydra front panel.
Содержание 2620A
Страница 4: ......
Страница 18: ...HYDRA Service Manual 1 2...
Страница 48: ...HYDRA Service Manual 1 32...
Страница 126: ...HYDRA Service Manual 3 2 3 30 Install the Handle and Mounting Brackets 3 15 3 31 Install the Instrument Case 3 15...
Страница 134: ...HYDRA Service Manual 3 10 Q Y U P J O A D B C V W G K L M E T X H I s25f eps Figure 3 5 2635A Assembly Details...
Страница 140: ...HYDRA Service Manual 3 16...
Страница 172: ...HYDRA Service Manual 5 2 5 28 Failure to Detect Memory PCA 5 29 5 29 Failure to Store Data 5 29...
Страница 234: ...HYDRA Service Manual 5A 34...
Страница 236: ...HYDRA Service Manual 6 2...
Страница 244: ...HYDRA Service Manual 6 10 A4 MP3 A4 H65 MP2 MP66 MP67 2620A 100 s58f eps Figure 6 1 2620A 2625A Final Assembly cont...
Страница 250: ...HYDRA Service Manual 6 16 A4 MP3 A4 H65 MP2 MP66 MP67 2620A 100 s62f eps Figure 6 2 2635A Final Assembly cont...
Страница 254: ...HYDRA Service Manual 6 20 2620A 1601 s63f eps Figure 6 3 2620A 2625A A1 Main PCA...
Страница 258: ...HYDRA Service Manual 6 24 2635A 1601 s64f eps Figure 6 4 2635A A1 Main PCA...
Страница 260: ...HYDRA Service Manual 6 26 2620A 4002 CKT 1 CKT 2 s65f eps Figure 6 5 A2 Display PCA...
Страница 265: ...List of Replaceable Parts Service Centers 6 6 31 2620A 1604 s67f eps Figure 6 7 A4 Analog Input PCA...
Страница 267: ...List of Replaceable Parts Service Centers 6 6 33 2620A 1605 s68f eps Figure 6 8 A5 IEEE 488 Interface PCA Option 05...
Страница 269: ...List of Replaceable Parts Service Centers 6 6 35 2625A 1606 s69f eps Figure 6 9 2625A A6 Memory PCA...
Страница 271: ...List of Replaceable Parts Service Centers 6 6 37 2635A 1606 s70f eps Figure 6 10 2635A A6 Memory Card I F PCA...
Страница 272: ...HYDRA Service Manual 6 38...
Страница 274: ...HYDRA Service Manual 7 2...
Страница 282: ...HYDRA Service Manual 7 10...
Страница 284: ...HYDRA Service Manual 8 2...
Страница 285: ...Schematic Diagrams 8 8 3 Figure 8 1 A1 Main PCA 2620A 2625A 2620A 1601 s88f eps...
Страница 288: ...Schematic Diagrams 8 8 6 Figure 8 1 A1 Main PCA 2620A 2625A cont 2620A 1001 3 of 4 s73f eps...
Страница 292: ...Schematic Diagrams 8 8 10 2635A 1001 2 of 5 Figure 8 2 A1 Main PCA 2635A cont s76f eps...
Страница 293: ...Schematic Diagrams 8 8 11 2635A 1001 3 of 5 Figure 8 2 A1 Main PCA 2635A cont s77f eps...
Страница 296: ...Schematic Diagrams 8 8 14 2620A 4002 CKT 1 CKT 2 Figure 8 3 A2 Display PCA s90f eps...
Страница 300: ...Schematic Diagrams 8 8 18 2620A 1003 2 of 3 Figure 8 4 A3 A D Converter PCA cont s82c eps...
Страница 301: ...Schematic Diagrams 8 8 19 2620A 1003 3 of 3 Figure 8 4 A3 A D Converter PCA cont s83c eps...
Страница 302: ...Schematic Diagrams 8 8 20 2620A 1604 Figure 8 5 A4 Analog Input PCA s92f eps...
Страница 304: ...Schematic Diagrams 8 8 22 2620A 1605 Figure 8 6 A5 IEEE 488 Interface PCA 2620A Only s93f eps...
Страница 306: ...Schematic Diagrams 8 8 24 2625A 1606 Figure 8 7 A6 Memory PCA 2625A s94f eps...
Страница 310: ...Schematic Diagrams 8 8 28...