![Fluke 2620A Скачать руководство пользователя страница 127](http://html1.mh-extra.com/html/fluke/2620a/2620a_manual_2309537127.webp)
General Maintenance
Introduction
3
3-3
3-1. Introduction
This section provides handling, cleaning, fuse replacement, disassembly, and assembly
instructions.
3-2. Warranty Repairs and Shipping
If your instrument is under warranty, see the warranty information at the front of this
manual for instructions on returning the unit. The list of authorized service facilities is
included in Section 6.
3-3. General
Maintenance
3-4. Required
Equipment
Equipment required for calibration, troubleshooting, and repair of the instrument is listed
in Section 4 (Table 4-1.)
Refer to the Fluke "Surface Mount Device Soldering Kit" for a list of special tools
required to perform circuit assembly repair. (In the USA, call 1-800-526-4731 to order).
3-5.
Power Requirements
Warning
To avoid shock hazard, connect the instrument powercord to a
power receptacle with earth ground.
If you have not already done so, plug the line cord into the connector on the rear of the
instrument. The instrument operates on any line voltage between 90V ac and 264V ac
and at any frequency between 45 and 440 Hz. However, the instrument is warranted only
to meet published specifications at 50/60 Hz. The instrument also operates from dc
power (9 to 16V dc). DC input power is connected to the rear input connector J6, pin 8
(DCH), and pin 7 (DCL). If both ac and dc power sources are connected to the
instrument, the ac power source is used if the ac line voltage exceeds approximately 8.3
times the dc voltage. Automatic switchover between ac and dc occurs without
interrupting instrument operation. The instrument draws a maximum of 10 VA on ac line
power or 4W on dc power.
3-6.
Static Safe Handling
All integrated circuits, including surface mounted ICs, are susceptible to damage from
electrostatic discharge (ESD). Modern integrated circuit assemblies are more susceptible
to damage from ESD than ever before. Integrated circuits today can be built with circuit
lines less than one micron thick, allowing more than a million transistors on a 1/4-inch
square chip. These submicron structures are sensitive to static voltages under 100 volts.
This much voltage can be generated on a dry day by simply moving your arm. A person
can develop a charge of 2,000 volts by walking across a vinyl tile floor, and polyester
clothing can easily generate 5,000 to 15,000 volts during movement against the wearer.
These low voltage static problems are often undetected because a static charge must be
in the 30,000 to 40,000 volt range before a person feels a shock.
Содержание 2620A
Страница 4: ......
Страница 18: ...HYDRA Service Manual 1 2...
Страница 48: ...HYDRA Service Manual 1 32...
Страница 126: ...HYDRA Service Manual 3 2 3 30 Install the Handle and Mounting Brackets 3 15 3 31 Install the Instrument Case 3 15...
Страница 134: ...HYDRA Service Manual 3 10 Q Y U P J O A D B C V W G K L M E T X H I s25f eps Figure 3 5 2635A Assembly Details...
Страница 140: ...HYDRA Service Manual 3 16...
Страница 172: ...HYDRA Service Manual 5 2 5 28 Failure to Detect Memory PCA 5 29 5 29 Failure to Store Data 5 29...
Страница 234: ...HYDRA Service Manual 5A 34...
Страница 236: ...HYDRA Service Manual 6 2...
Страница 244: ...HYDRA Service Manual 6 10 A4 MP3 A4 H65 MP2 MP66 MP67 2620A 100 s58f eps Figure 6 1 2620A 2625A Final Assembly cont...
Страница 250: ...HYDRA Service Manual 6 16 A4 MP3 A4 H65 MP2 MP66 MP67 2620A 100 s62f eps Figure 6 2 2635A Final Assembly cont...
Страница 254: ...HYDRA Service Manual 6 20 2620A 1601 s63f eps Figure 6 3 2620A 2625A A1 Main PCA...
Страница 258: ...HYDRA Service Manual 6 24 2635A 1601 s64f eps Figure 6 4 2635A A1 Main PCA...
Страница 260: ...HYDRA Service Manual 6 26 2620A 4002 CKT 1 CKT 2 s65f eps Figure 6 5 A2 Display PCA...
Страница 265: ...List of Replaceable Parts Service Centers 6 6 31 2620A 1604 s67f eps Figure 6 7 A4 Analog Input PCA...
Страница 267: ...List of Replaceable Parts Service Centers 6 6 33 2620A 1605 s68f eps Figure 6 8 A5 IEEE 488 Interface PCA Option 05...
Страница 269: ...List of Replaceable Parts Service Centers 6 6 35 2625A 1606 s69f eps Figure 6 9 2625A A6 Memory PCA...
Страница 271: ...List of Replaceable Parts Service Centers 6 6 37 2635A 1606 s70f eps Figure 6 10 2635A A6 Memory Card I F PCA...
Страница 272: ...HYDRA Service Manual 6 38...
Страница 274: ...HYDRA Service Manual 7 2...
Страница 282: ...HYDRA Service Manual 7 10...
Страница 284: ...HYDRA Service Manual 8 2...
Страница 285: ...Schematic Diagrams 8 8 3 Figure 8 1 A1 Main PCA 2620A 2625A 2620A 1601 s88f eps...
Страница 288: ...Schematic Diagrams 8 8 6 Figure 8 1 A1 Main PCA 2620A 2625A cont 2620A 1001 3 of 4 s73f eps...
Страница 292: ...Schematic Diagrams 8 8 10 2635A 1001 2 of 5 Figure 8 2 A1 Main PCA 2635A cont s76f eps...
Страница 293: ...Schematic Diagrams 8 8 11 2635A 1001 3 of 5 Figure 8 2 A1 Main PCA 2635A cont s77f eps...
Страница 296: ...Schematic Diagrams 8 8 14 2620A 4002 CKT 1 CKT 2 Figure 8 3 A2 Display PCA s90f eps...
Страница 300: ...Schematic Diagrams 8 8 18 2620A 1003 2 of 3 Figure 8 4 A3 A D Converter PCA cont s82c eps...
Страница 301: ...Schematic Diagrams 8 8 19 2620A 1003 3 of 3 Figure 8 4 A3 A D Converter PCA cont s83c eps...
Страница 302: ...Schematic Diagrams 8 8 20 2620A 1604 Figure 8 5 A4 Analog Input PCA s92f eps...
Страница 304: ...Schematic Diagrams 8 8 22 2620A 1605 Figure 8 6 A5 IEEE 488 Interface PCA 2620A Only s93f eps...
Страница 306: ...Schematic Diagrams 8 8 24 2625A 1606 Figure 8 7 A6 Memory PCA 2625A s94f eps...
Страница 310: ...Schematic Diagrams 8 8 28...