![Fluke 1577 Скачать руководство пользователя страница 24](http://html1.mh-extra.com/html/fluke/1577/1577_calibration-manual_2309485024.webp)
1587/1577
Calibration
14
3.
Remove the old glass LCD display from the bezel.
Note
Make sure the new LCD display is clean (free of lint and finger prints)
before placing it in position in the bezel.
4.
Position the new glass LCD display in the bezel; the silver face should face the
rear, and the stepped portion of the glass should be directly under the elastomeric
slot on the bezel.
5.
Drop the elastomeric strip into its slot on the bezel.
6.
Position the translucent light-dispersing back panel over the silver side of the
glass LCD display. Make sure the guide pins on the light dispersing back panel
are facing up and that they are on the side opposite the elastomeric strip.
7.
While holding the LCD assembly (face down) in one hand, position the PCA
(fuse side up) over the bezel; match the guide holes in the PCA with the plastic
guide pins on the bezel and the translucent light dispersing back panel. After the
LCD assembly is in position, lock it in place by pressing (below the display) the
bezel against the PCA; listen for the tab on the bezel to snap (lock) into position
on the PCA.
8.
Secure the PCA to the LCD assembly by installing the two screws that attach the
PCA to the bezel. These screws also ensure contact between the glass LCD
display, the elastomeric strip, and the PCA.
Light-dispersing
Back Panel
Glass LCD
Display
PCA
Elastomeric
Strip
Plastic
Bezel
Screws
Gray Plastic
Tab
ebp05f.eps
Figure 6. Accessing the LCD