5.2.1.12
Thermocouple Polynomial Conversion
EDIT TCS 03
PROBE SER#: 1
CONVERSION: TC-POLY
CJC: INTERNAL
C0: 0.0
CJC TEMP: 0.0
C1: 0.038562
$
NOTE:
An Application Note for use of Tungsten-Rhenium and other
thermocouples is available at www.hartscientific.com.
The following equation is used for the thermocouple polynomial conversion:
E t
c t
i
i
i
( [
])[
]
°
=
=
∑
C
V
μ
1
15
NOTE:
The EMF (E) in the above equation is in microvolts. If the cali-
bration equation for the coefficients of the probe being used is in milli-
volts, each coefficient needs to be multiplied by 1000 before entering the
coefficients into the instrument.
The user-defined parameters for the thermocouple polynomial conversion are
the coefficients C0 (
c
0
) through C15 (
c
15
). The user can specify internal or ex-
ternal cold-junction compensation (CJC). If external CJC is selected, the user
can enter the temperature of the CJC reference. DC1 and DC2 are only accessi-
ble when type R is used.
5.2.2
Copy Probe
The
COPY PROBE
soft-key can be used to copy an entire set of probe param-
eters from one input channel to another. The parameters are copied in two
steps—select the source channel then select the destination channel. A window
appears requesting you to select the source channel. Each channel in the list is
identified with the device name, the channel number, and the probe serial num-
ber. The list includes two
USER
channels that can be used for temporary stor-
age. The source channel is selected using the
UD
buttons and pressing
Enter
.
43
5 Soft-Key Functions
Probe Menu
Содержание 1560 Black Stack
Страница 1: ...1560 Black Stack Thermometer Readout User s Guide Rev 932001 Hart Scientific...
Страница 20: ...1560 Thermometer Readout User s Guide 8 Figure 2 System Diagram...
Страница 27: ...15 2 Introduction Data Processing Figure 5 Data Flow...
Страница 70: ...RESET SYSTEM PARAMETERS PRESS ENTER TO CONTINUE PRESS EXIT TO CANCEL 61 5 Soft Key Functions System Menu...