background image

116C 

Users Manual 

Battery Saver ("Sleep Mode") 

The Meter automatically enters "Sleep mode" and blanks 
the display if there is no function change, range change or 
button press for 20 minutes. Pressing any button or 
turning the rotary switch awakens the Meter. To disable the 
Sleep mode, hold down the 

g

 

button while turning the 

Meter on. The Sleep mode is always disabled in the MIN 
MAX AVG mode. 

MIN MAX AVG Recording Mode 

The MIN MAX AVG recording mode captures the minimum 
and maximum input values (ignoring overloads), and 
calculates a running average of all readings. When a new 
high or low is detected, the Meter beeps.  

 

Put the Meter in the desired measurement function and 
range. 

 

Press 

p

 to enter MIN MAX AVG mode. 

 

M

and MAX are displayed and the highest 

reading detected since entering MIN MAX AVG is 
displayed. 

 

Press 

p

 to step through the low (MIN), average 

(AVG), and present readings. 

 

To pause MIN MAX AVG recording without erasing 
stored values, press 

f

K

 is displayed. 

 

To resume MIN MAX AVG recording, press 

f

again. 

 

To exit and erase stored readings, press 

p

 for at 

least one second or turn the rotary switch.  

Display HOLD  

XW

Warning 

To avoid electric shock, when Display 
HOLD is activated, be aware that the 
display will not change when you 
apply a different voltage. 

In the Display HOLD mode, the Meter freezes the display. 

1.

 

Press 

f

to activate Display HOLD. (

K

 is 

displayed.)  

2.

 

To exit and return to normal operation, press 

f

 or 

turn the rotary switch. 

Backlight 

Press 

Q

 to toggle the backlight on and off. The backlight 

automatically turns off after 40 seconds. To disable 
backlight auto-off, hold down 

Q

 while turning the Meter 

on. 

Manual and Autoranging 

The Meter has both Manual and Autorange modes. 

 

In the Autorange mode, the Meter selects the range with 
the best resolution. 

 

In the Manual Range mode, you override Autorange and 
select the range yourself. 

Содержание 116C

Страница 1: ...PN 2538707 July 2006 Rev 1 2 07 2006 2007 Fluke Corporation All rights reserved Printed in China All product names are trademarks of their respective companies 116C True rms Multimeter Users Manual...

Страница 2: ...return authorization information then send the product to that Service Center with a description of the problem THIS WARRANTY IS YOUR ONLY REMEDY NO OTHER WARRANTIES SUCH AS FITNESS FOR A PARTICULAR P...

Страница 3: ...t against transients in fixed equipment installations at the distribution level Contacting Fluke To contact Fluke call USA 1 888 99 FLUKE 1 888 993 5853 China 010 65123435 Europe 31 402 675 200 Japan...

Страница 4: ...measurements Verify the Meter s operation by measuring a known voltage If in doubt have the Meter serviced Do not apply more than the rated voltage as marked on Meter between terminals or between any...

Страница 5: ...ck the test leads for continuity before use Do not use if the readings are high or noisy Do not use the Auto Volts function to measure voltages in circuits that could be damaged by this function s low...

Страница 6: ...s No Symbol Meaning A s The Meter function is set to Continuity B R The Meter function is set to Diode Test C O Input is a negative value D Y X Unsafe voltage Measured input voltage 30 V or voltage ov...

Страница 7: ...Direct current or alternating current J N Battery low warning K 610000 mV Indicates the Meter s range selection L Bar graph Analog display M Auto Volts Auto Manual The Meter is in the Auto Volts funct...

Страница 8: ...apacitance frequency temperature microamps and testing diodes Error Messages bAtt Battery must be replaced before the Meter will operate CAL Err Calibration required Meter calibration is required befo...

Страница 9: ...001 V to 600 V l AC voltage from 6 0 to 600 mV dc coupled DC voltage from 0 1 to 600 mV Ohms from 0 1 to 40 M s Continuity beeper turns on at 20 and turns off at 250 y Temperature from 40 C to 400 C...

Страница 10: ...tep through the low MIN average AVG and present readings To pause MIN MAX AVG recording without erasing stored values press f K is displayed To resume MIN MAX AVG recording press fagain To exit and er...

Страница 11: ...Up Options To select a Power Up Option hold down the button indicated in the following table while turning the Meter on Power Up Options are canceled when you turn the Meter off and when sleep mode is...

Страница 12: ...t power and discharge all high voltage capacitors before testing resistance continuity diodes or capacitance Testing for Continuity eeo06f eps Note The continuity function works best as a fast conveni...

Страница 13: ...ac voltage measurement based on the input applied between the and COM jacks This function also sets the Meter s input impedance to approximately 3 k to reduce the possibility of false readings due to...

Страница 14: ...rement when the open circuit potential to earth is 600 V Use the proper switch position and range for your measurement Flame sensor probe Control Module eeo08f eps To measure flame rectification circu...

Страница 15: ...king Basic Measurements 13 Measuring Capacitance eeo05f eps Measuring Temperature 80BK Type K Thermocouple Probe Vent or Pipe RANGE eeo10f eps XW Warning To avoid risk of electric shock do NOT connect...

Страница 16: ...turn the frequency measurement function on and off Frequency works with ac volts only 2 In frequency the bar graph and range annunciator indicate the AC voltage present 3 Select progressively lower ra...

Страница 17: ...eps Measuring Current above 600 A The millivolt and voltage function of the Meter can be used with an optional mV A output Current Probe to measure currents that exceed the rating of the Meter Make su...

Страница 18: ...nd temperature In frequency the bar graph and range annunciator indicates the underlying voltage up to 1 kHz The number of segments indicates the measured value and is relative to the full scale value...

Страница 19: ...o lift the door slightly 4 Lift the door straight up to separate it from the case The battery fits inside the battery door which is then inserted into the case bottom edge first until it is fully seat...

Страница 20: ...gital 6 000 counts updates 4 sec Bar Graph 33 segments updates 32 sec Temperature Operating 10 C to 50 C Storage 40 C to 60 C Temperature Coefficient 0 1 x specified accuracy C 18 C or 28 C Operating...

Страница 21: ...of Reading Counts DC millivolts 600 0 mV 0 1 mV 0 5 2 DC Volts 6 000 V 60 00 V 600 0 V 0 001 V 0 01 V 0 1 V 0 5 2 DC 45 to 500 Hz 500 Hz to 1 kHz Auto V LoZ 1 True rms 600 0 V 0 1 V 2 0 3 4 0 3 45 to...

Страница 22: ...600 0 6 000 k 60 00 k 600 0 k 6 000 M 40 00 M 0 1 0 001 k 0 01 k 0 1 k 0 001 M 0 01 M 0 9 2 0 9 1 0 9 1 0 9 1 0 9 1 5 2 Diode test 2 000 V 0 001 V 0 9 2 Capacitance 1000 nF 10 00 F 100 0 F 9999 F 1 nF...

Страница 23: ...ecified from 1 to 100 of range Auto V LoZ is specified from 0 V Because inputs below 1 of range are not specified it is normal for this and other true rms meters to display non zero readings when the...

Страница 24: ...al Mode Rejection Volts AC 5 M 100 pF 60 dB at dc 50 or 60 Hz Volts DC 10 M 100 pF 100 dB at dc 50 or 60 Hz 60 dB at 50 or 60 Hz Auto V LoZ 3 k 500 pF 60 dB at dc 50 or 60 Hz Open Circuit Test Voltage...

Отзывы: