8588A/8558A
Remote Programmer's Manual
36
Table 6. Command Summary (cont.)
Keyword
Description
:RANGe? [MIN|MAX|DEF]
Returns the selected range, or if specified, the MIN,
MAX, or Default range
:RESolution <
nrf
>|MIN|MAX|DEF
Set maximum expected value or min, max or default
resolution; for example, range is 1V, Resolution
<nrf>
. = 0.0001 (100 µV), the measurement is
returned with a resolution of +1.000E-4
:RESolution? [MIN|MAX|DEF]
Returns the selected resolution, or if specified, the
minimum, maximum or default range
[:SENSe]:XCURrent:AC
:COUNter:BLIMit[:STATe]
<boolean>
Turn the frequency path bandwidth limit ON or OFF
:COUNter:BLIMit[:STATe]
?
Return the state of the frequency path bandwidth limit
setting. Returns 1 for Auto range ON, 0 for auto
range OFF.
:COUNter:COUPling AC | DC
Set the frequency path coupling to ac or dc. Note that
if the signal path coupling is ac, frequency path
coupling is forced to ac.
:COUNter:COUPling?
Return the frequency path coupling setting
:COUNter:GATE:AUTO[:STATe]
<boolean>
Turn Auto selection of counter gate ON or OFF
:COUNter:GATE:AUTO[:STATe]?
Return the state of the counter gate auto setting, 1 for
Auto gate ON, 0 for auto gate OFF
:COUNter:GATE <
nrf
>|MIN|MAX|DEF
Sets the secondary frequency counter measurement
gate
:COUNter:GATE? [MIN|MAX|DEF]
Return counter gate setting, or if specified, the MIN,
MAX, or Default setting
:COUPling AC|DC
Set the ac current signal path coupling to ac or dc.
Note that if the signal path coupling is ac, frequency
path coupling is forced to ac.
:COUPling?
Return the frequency path coupling setting
:FILTer <
nrf>
|MIN|MAX|DEF
Sets the input filter to
value
or to the minimum,
maximum or default setting. The filter setting range is
0.1Hz to 1kHz, the discrete settings are: 0.1 Hz,
1 Hz, 10 Hz, 40 Hz, 100 Hz, 1 kHz. The discrete
setting closest to <nrf> will be selected.
:FILTer? [MIN|MAX|DEF]
Return the filter setting