Table 6. Command Summary (cont.)
Keyword
Description
:COUNter:GATE <range>|MIN|MAX|DEF
Sets the secondary frequency counter measurement gate
:COUNter:GATE? [MIN|MAX|DEF]
Returns the secondary frequency counter measurement gate
:COUPling:SIGNal
AC1M|AC10m|DC1M|DC10m|DCAuto
Sets the coupling and input impedance of the signal path
:COUPling:SIGNal?
Returns the signal path coupling and bandwidth setting
:FILTer <
nrf>
|MIN|MAX|DEF
Sets the input filter to
value
or to the minimum, maximum or
default setting. The filter setting range is 0.1 Hz to 1 kHz, the
discrete settings are: 0.1 Hz, 1 Hz, 10 Hz, 40 Hz, 100 Hz, 1 kHz.
The discrete setting closest to <nrf> will be selected.
:FILTer? [MIN|MAX|DEF]
Returns the Filter setting, or if specified, the MIN, MAX, or
Default setting
:RANGe:AUTO[:STATe]
<boolean>
Turns Auto range ON or OFF
:RANGe:AUTO[:STATe]?
Returns 1 for Auto range ON, 0 for auto range OFF
:RANGe <nrf>|MIN|MAX|DEF
Set maximum expected value or MIN, MAX, or Default range
:RANGe? [MIN|MAX|DEF]
Returns the selected range or, or if specified, the MIN, MAX, or
Default range
:RESolution <nrf>|MIN|MAX|DEF
Set maximum expected value or min, max or default resolution;
for example, range is 1 V, Resolution
<nrf>
. = 0.0001 (100 µV),
the measurement is returned with a resolution of +1.000E-4
:RESolution? [MIN|MAX|DEF]
Returns the selected resolution, or if specified, the MIN, MAX, or
Default range
:SECondary
OFF|FREQuency|PERiod|PTP
|PPEak|NPEak|CFACtor
Sets the secondary reading type
OFF = Secondary reading is not shown
FREQuency = Secondary reading is frequency
PERiod = Secondary reading is period
PTP = Secondary reading is peak-to-peak
PPEak = Secondary reading is positive peak
NPEak = Secondary reading is negative peak
CFACtor = Secondary reading is the crest factor