Controller Maintenance
Maintenance Procedures
4
4-5
Pressure Leakage Test
Leak Test is a function of the Controller that checks the Controller and UUT connection
for pressure leaks. To detect leaks, the Controller generates pressure and monitors the
rate of pressure change over a specified time period. The total change in pressure (dP)
and average rate of change is shown on the Leak Test menu and stays on the menu until
the test is reaccomplished. See Table 4-3 for a list of leak test parameters.
To test for a leak:
1.
Connect the male end of the test hose to the female
TEST
port on the Controller.
2.
Connect female end of the test hose to the male
QC adapter
on the UUT.
Note
A leak test can be done without a UUT connected to troubleshoot the
Controller.
3.
Navigate to and select
on the bottom of the Main menu. The Leak Test
menu opens.
4.
Enter the leak test time in seconds into the
Leak Time
field.
Note
The leak test menu contains an Average Rate, dP (pressure change),
and Time values. These values are leak test result values and are not
adjustable settings. When the menu opens, the values shown are from
the last test completed and will update after the leak test is complete.
5.
Navigate to and select
to start the leak test.
6.
If the leak rates are excessive, search for a leak path around the adapter fittings first.
To search for leaks, use a spray bottle with a mixture of water and diluted soap to
spray on the fittings. If a there is a leak, bubbles from the soap visually show where
the leak is. See Figure 4-1.
Table 4-3. Leak Test Parameters
Parameter Description
Leak Time
Amount of time to run the leak test.
Average Rate
Shows the average rate of change of pressure for the test completed.
Measured in units for each second.
dP
Shows the total change in pressure for the test completed.
Time
Shows the length of time the test ran.
Содержание 4322-SYS-NAVAIR
Страница 1: ...4322 SYS NAVAIR Automated Pressure Calibrator Service Manual...
Страница 10: ...4322 Service Manual vi...
Страница 14: ...4322 Service Manual 1 2...
Страница 22: ...4322 Service Manual 1 10...
Страница 24: ...4322 Service Manual 2 2...
Страница 30: ...4322 Service Manual 2 8 hbx064 eps Figure 2 2 Pressure Valve Schematic...
Страница 42: ...4322 Service Manual 2 20...
Страница 44: ...4322 Service Manual 3 2...
Страница 54: ...4322 Service Manual 4 2...
Страница 94: ...4322 Service Manual 4 42...
Страница 96: ......
Страница 104: ...4322 Service Manual 5 10 Intensifier Controller Fuse Holder 1 2 3 4 A B C gxj081 eps Figure 5 1 Fuse Replacement...
Страница 106: ...4322 Service Manual 5 12 1 2 3 hbx008 eps Figure 5 2 Intensifier Case Removal...
Страница 132: ...4322 Service Manual 6 2...
Страница 138: ...4322 Service Manual 7 2...
Страница 164: ...4322 Service Manual 7 28 H1 H2 H10 H3 H6 H9 H4 H8 H6 A1 H7 H6 H5 A2 A3 hbx043 eps Figure 7 13 A2 Booster Pump Assembly...
Страница 170: ...4322 Service Manual 7 34...
Страница 172: ...4322 Service Manual 8 2...
Страница 180: ...4322 Service Manual 9 2...
Страница 186: ...4322 Service Manual 9 8...