P3000 Series
Users Manual
1-4
Temperature and Air Density variations are less significant than gravity.
Variations should be corrected for when maximum accuracy is required.
Temperature variation example:
Deadweight Tester calibrated temperature
20 °C
Operating temperature
24 °C
Percentage change per °C
0.002%
Indicated Pressure
250 psi
(
)
250
100
002
.
0
24
20
250
x
x
PRESSURE
TRUE
−
+
=
psi
PRESSURE
TRUE
98
.
249
=
Model Variations
This manual covers many pressure ranges and models within the P3000 series of
instruments.
The following illustrations detail the physical appearance of the six basic models:
Model 3015
gjn024.bmp
Figure 1-1. Single PCU Instrument