Demo Instructions
9
Warning: defibrillators/external pacemakers are capable of emitting lethal
high voltages/energies and should be operated only by trained professionals.
THE CHARGE TIME TEST
Tests the batteries and charging capacity in the defibrillator
1. Select F3 for CHARGE TIME test.
2. Select F3 for MEASURE
a.
Notice CHARGE TIME screen counting down “5-4-3-2-1”
3. Select desired energy level and then charge defibrillator.rillator when the count reaches
zero and the internal beeper sounds... When the selected energy level is reached,
discharge the paddles.
a.
Notice the results screen “CHARGE TIME = ___ s”.
Please note that typical
charge times are under 10 seconds.