Demo Instructions

 

 

 

Warning: defibrillators/external pacemakers are capable of emitting lethal 

high voltages/energies and should be operated only by trained professionals. 

 

 

 
THE SYNCHRONIZATION  TEST (Cardioversion)

 

Measures the time delay in milliseconds (ms) between the peak of the R wave and the 
discharging of the defibrillator pulse.  A wide range of simulated ECG waveforms are 
available in the analyzer for performing this test.  Atrial fibrillation (AFIB) is the most 
representative clinical waveform for demonstration purposes. Two selections are 
available; AFIB-Coarse is the default simulation.  

 
 
 
1. Press 

F2“Sync”

   

 

 

 
2. Press 

F1 “Set Wave”

 Use UP/DOWN arrow keys to select an ECG waveform.  AFIB 

is chosen for the purpose of this demonstration.  
 

 

 
3. Note ability to change AMPLITUDE and RATE.  Select F5 for DONE.   
 
4. Select F2 to select WAVE CONVERSION.  Three choices are listed: CONVERT, NO 
CONVERT and ASYSTOLE.  Convert is chosen for the purpose of this demonstration 
 

 

 

Содержание IMPULSE 6000D

Страница 1: ...eart rate setting facilitates rate meter alarm testing 1 BPM per step 7 Easy to use connection posts for 12 Lead ECG 10 electrodes 8 Wide range of Arrhythmia selections 9 Auto range defib pacer measur...

Страница 2: ...d professionals 3 Future upgradeable firmware software and new waveforms 4 Global support network Low level ECG Power switch Navigator Function Pacemaker input 50 to 1500 ohms t LCD 1 Plug in defib pa...

Страница 3: ...hbutton The main menu appears on the analyzer s display The battery ICON in the upper right of the display indicates the relative charge level of the analyzer s internal battery If ICON is showing les...

Страница 4: ...capable of emitting lethal high voltages energies and should be operated only by trained professionals 3 To set up unit for test apply ECG leads to the ten AHA color on the left side of the post and...

Страница 5: ...B mode on the left panel then select a test a Press F1 ENERGY b Press F1 Set Wave and then select VFib from list using the UP DOWN keys c Press F2 Wave Conversion and select convert from list using th...

Страница 6: ...onals 3 The analyzer calculates the energy delivered over the load resistance and displays the result in joules under the result Press F4 for summary 4 Show on the monitor that ECG waveform is convert...

Страница 7: ...are available in the analyzer for performing this test Atrial fibrillation AFIB is the most representative clinical waveform for demonstration purposes Two selections are available AFIB Coarse is the...

Страница 8: ...be operated only by trained professionals 5 Power on the defibrillator Select the desired energy place the paddles and discharge Note for the purpose of this demonstration AFIB is converted to a norma...

Страница 9: ...ty in the defibrillator 1 Select F3 for CHARGE TIME test 2 Select F3 for MEASURE a Notice CHARGE TIME screen counting down 5 4 3 2 1 3 Select desired energy level and then charge defibrillator rillato...

Страница 10: ...d and sensitivity tests measures and displays refractory periods etc 1 Connect pacer to the analyzer through either defibrillator input 50 ohms only or pacer input 50 to 1500 ohms 2 Select PACER mode...

Страница 11: ...Turn pacer on and adjust pacer rate and output 8 Observe pacer Measurement result screen 9 Select F4 to view more tests F1 SENSITIVITY F2 REFRACTORY PERIOD Sensitivity test Demand sensitivity is the...

Страница 12: ...MANCE can choose different wave forms such as square pulse sine and triangle c F3 PACER INTERACTIVE Selections Under F3 are a Non capture the analyzer operates in continuous mode However every 10th pa...

Страница 13: ...e operated only by trained professionals b F2 PREMATURE c F3 VENTRICULAR d F4 MORE b Selects F4 again a Wave types i F1 CONDUCTION ii F2 TV PACED iii F3 R WAVE DETECTION iv Noise Immunity For more inf...