Detailed Specifications
Accuracy
Two Terminal Mode
Test current >200 mA ac into 500 m
Ω
........
±
(2 % of r 0.015
Ω
) for 0.0 to 2.0
Ω
Test current 1-16 A ac .................................
±
(2 % of r 0.015
Ω
) for 0.0 to 0.2
Ω
±
(5 % of r 0.015
Ω
) for 0.2 to 2.0
Ω
Four Terminal Mode
Test current >200 mA ac into 500 m
Ω
........
±
(2 % of r 0.005
Ω
) for 0.0 to 2.0
Ω
Test current 1-16 A ac .................................
±
(2 % of r 0.005
Ω
) for 0.0 to 0.2
Ω
±
(5 % of r 0.005
Ω
) for 0.2 to 2.0
Ω
Additional error caused by series inductance
Resistance
Series Inductance
0 µH
100 µH
200 µH
400 µH
0.000
Ω
0.000
Ω
0.030
Ω
0.040
Ω
0.050
Ω
0.020
Ω
0.000
Ω
0.025
Ω
0.030
Ω
0.040
Ω
0.040
Ω
0.000
Ω
0.020
Ω
0.025
Ω
0.030
Ω
0.060
Ω
0.000
Ω
0.015
Ω
0.020
Ω
0.025
Ω
0.080
Ω
0.000
Ω
0.010
Ω
0.015
Ω
0.020
Ω
0.100
Ω
0.000
Ω
0.010
Ω
0.010
Ω
0.015
Ω
>0.100
Ω
0.000
Ω
0.010
Ω
0.010
Ω
0.010
Ω