Fluigent PRESSURE UNIT Скачать руководство пользователя страница 7

7

SPECIFICATIONS

Software compatibility:

 The 

PRESSURE UNIT

 sensor is compatible with 

Fluigent 

All-in-one

 software to display the value measured in real time, 

or with the 

Software Development Kit

 to output the measured value in 

custom software application. Download those softwares using the 

Fluigent 

Software Installer

 downloadable from the website, or provided 

USB stick

.

Technical specifications

Measurement 

sampling

40 ms

40 ms

40 ms

Internal volume

22 µL

22 µL

22 µL

Compensated 

temperature range

0 to 50°C

0 to 50°C

0 to 50°C

Connection fittings

1/4’’-28 Flat 

bottom

1/4’’-28 Flat 

bottom

1/4’’-28 Flat bottom

Recommended tubing

1/16’’ OD

1/16’’ OD

1/16’’ OD

Dimensions

50 x 30 x 20 mm

50 x 30 x 20 mm

50 x 30 x 20 mm

Material

PEEK, EPDM, 

Silicium

PEEK, EPDM, 

Silicium

PEEK, EPDM, 

Silicium

Maximum operating 

altitude

Up to 2000 m

Up to 2000 m

Up to 2000 m

Maximum relative 

humidity

80% (0°C to 

31°C) 

50% (until 50°C)

80% (0°C to 31°C) 

50% (until 50°C)

80% (0°C to 31°C) 

50% (until 50°C)

Electrical specifications

Voltage range 

(Direct current)

5V 

5 V   

5 V

Maximal power

10 mW

10 mW

10 mW

Содержание PRESSURE UNIT

Страница 1: ...USER S MANUAL PRESSURE UNIT INLINE PRESSURE SENSOR M...

Страница 2: ...s a perfusion system which automates perfusion or timed injection protocols It allows for the sequential delivery of up to 10 different solutions at the desired flow rate into a microfluidic chip perf...

Страница 3: ...QUICK START GUIDE 4 PRESSURE UNIT USER S MANUAL PRODUCT OVERVIEW 6 SPECIFICATIONS 7 WARRANTY TERMS 9 TECHNICAL SUPPORT 10...

Страница 4: ...g the provided connectics and tubing kit provided by Fluigent connect the PRESSURE UNIT into the fluidic path Note The fluid can flow following the direction of the arrow indicated on the PRESSURE UNI...

Страница 5: ...lers using LineUpTM local control or using Fluigent dedicated software on a computer and get the measured pressure in the fluidic path by the PRESSURE UNIT directly on Fluigent All in One software int...

Страница 6: ...bedirectlypluggedtoaPCwithUSBconnectionanddisplayin livethemeasurementonFluigentAll in Onesoftwareinterface Theuser s canoutputthisvalueforcustomsoftwareapplicationusingtheSoftware Development Kit 5 2...

Страница 7: ...ms Internal volume 22 L 22 L 22 L Compensated temperature range 0 to 50 C 0 to 50 C 0 to 50 C Connection fittings 1 4 28 Flat bottom 1 4 28 Flat bottom 1 4 28 Flat bottom Recommended tubing 1 16 OD 1...

Страница 8: ...Accuracy mean of max range 2 to 3 mbar 0 6 typ to 0 9 10 to 20 mbar 1 0 typ to 2 0 16 to 40 mbar 0 3 typ to 0 6 Zero shift 6 9 mbar 2 span 10 mbar 1 span 70 mbar 1 span Repeatability Hysteresis 1 4 mb...

Страница 9: ...time for Fluigent service representative to discuss and find a solution to fix the issue Repairs will be made remotely whenever possible If more action is needed the system will need to be sent back t...

Страница 10: ...your local office For tutorial videos about Fluigent products please visit Fluigent on YouTube Fluigent Fluigent S A S 33 1 77 01 82 65 Fluigent Inc 1 978 934 5283 Fluigent GmbH 49 3641 277 652 For a...

Страница 11: ...11 NOTES...

Страница 12: ...VERSION FEB 2021...

Отзывы: