5.3
TERMINAL CONNECTORS INTRINSICALLY SAFE APPLICATIONS
Terminal connectors F011-P-(PC / PD / PX)-XI-(ZB):
Fig. 31: Overview terminal connectors Type XI - Intrinsically Safe applications
•
Type PC:
offers - additional to type PX - an internal Intrinsically Safe lithium battery. This ATEX
certified battery (FW-LiBAT-xxx) may be changed in hazardous area.
•
Type PD:
offers with terminal 6 - additional to type PX - a real sensor supply of 8.2V DC to power
a Namur sensor for example.
•
Type PX:
as standard, all intrinsically safe products are supplied with terminal 4 and 5 to power
the product externally.
5.4
CONFIGURATION EXAMPLES INTRINSICALLY SAFE APPLICATIONS
5.4.1
F011-P-PC-(PX)-XI-(ZB) BATTERY POWERED - Ex ia IIC/IIIC
Fig. 32: F011-P-PC-(PX)-XI-(ZB) - Ex ia IIC/IIIC - Intrinsically Safe application.
*
Sensor supply voltage for pulse type P: Terminal 3: 1.2-3.2V DC.
Please note: Type PX may be used in combination with the battery (Type PC).
PX will power the F011-P, the battery will be disabled automatically until the power is disconnected.
*Sensor supply voltage for pulse flowmeter type P : Terminal 3: 1.2 - 3.2V DC.
Please note: type PX may be used in combination with the battery (type PC).
PX will power the unit; the battery will be disabled automatically till power is disconnected.
HAZARDOUS AREA
SAFE AREA
I.S. flowmeter
input type: P
pulse
Configuration example IIA - IIB and IIC
battery powered unit
F010-P-PC-(PX)-XI-(ZB) -
Basic power supply type PX:
8-30V DC
(not used in this example).
Backlight option: type ZB
(not used in this example).
TERMINAL CONNECTORS
F0-series
1
2
3
Common ground
Signal
Supply *
C
irc
ui
t d
ep
en
ds
o
n
ty
pe
o
f s
ig
na
l
4
5
Common ground
Main supply
9
1
0
Common ground
Supply backlight