![Fluidwell E126-P-EL Скачать руководство пользователя страница 66](http://html1.mh-extra.com/html/fluidwell/e126-p-el/e126-p-el_manual_2308976066.webp)
User Manual
E126-P-EL
Page 66
FW_E126-P-EL_M_v0301-01_EN
MODBUS-REGISTERS: DISPLAY
[d] 58
[h] 0x03A
40059
bar graph
enable
1
R/W
uint16 0=disable
1=enable
[d] 59
[h] 0x03B
40060
bar graph range
2
R/W
uint32 0…9999999
Representation: unit, time, decimals depending on variables
48, 49, 50
MODBUS-REGISTERS: FLOWMETER
PDU
ADDRESS
HOLDING
REGISTER
VARIABLE
NO. OF
REG’S
R/W
TYPE VALUES / REPRESENTATION / REMARKS
[d] 96
[h] 0x060
40097
flowmeter signal
1
R/W
uint16 0=NPN
1=NPN LP
2=reed
3=reed LP
4=PNP
5=PNP LP
6=Namur
7=coil hi
8=coil lo
9=act 8.2V
10=act 24V
[d] 1051
[h] 0x41B
41052
K-factor unit
type
1
R/W
uint16 0=hand
1=volumetric
2=mass
[d] 1050
[h] 0x41A
41051
K-factor unit
1
R/W
uint16 0=none
1=L
2=m
3
3=US GAL
4=I GAL
5=CF
6=OilBBL
7=kg
8=ton
9=LB
10=US ton
[d] 1046
[h] 0x416
41047
K-factor
2
R/W
uint32 1…9999999
Representation: 0.000010…9999999 depending on variable
54: decimals K-factor
[d] 1049
[h] 0x419
41050
K-factor
decimals
1
R/W
uint16 0…6
MODBUS-REGISTERS: TEMPERATURE - WITH TYPE TA (4-20MA)
PDU
ADDRESS
HOLDING
REGISTER
VARIABLE
NO. OF
REG’S
R/W
TYPE VALUES / REPRESENTATION / REMARKS
[d] 528
[h] 0x210
40529
unit
1
R/W
uint16 0=K
1=°C
2=°F
[d] 530
[h] 0x212
40531
span
2
R/W
uint32 0.00001…999999 K
Representation: decimals depending on variable 533
[d] 533
[h] 0x215
40534
decimals span
1
R/W
uint16 0…5
[d] 530
[h] 0x212
40531
span
2
R/W
uint32 0.00001…999999 K
Representation: decimals depending on variable 533
[d] 183
[h] 0x0B7
40184
filter
1
R/W
uint16 1…99
[d] 186
[h] 0x0BA
40187
calib-low
1
R/W
uint16 0=cal set
1=calibrate
2=default
[d] 187
[h] 0x0BB
40188
calib-high
1
R/W
uint16 0=cal set
1=calibrate
2=default
MODBUS-REGISTERS: TEMPERATURE - WITH TYPE TP (PT100)
PDU
ADDRESS
HOLDING
REGISTER
VARIABLE
NO. OF
REG’S
R/W
TYPE VALUES / REPRESENTATION / REMARKS
[d] 209
[h] 0x0D1
40210
display
1
R/W
uint16 0=K
1=°C
2=°F
[d] 208
[h] 0x0D0
40209
number of wires
1
R/W
uint16 0=2
1=3
2=4
[d] 199
[h] 0x0C7
40200
filter
1
R/W
uint16 1…99
MODBUS-REGISTERS: FORMULA
PDU
ADDRESS
HOLDING
REGISTER
VARIABLE
NO. OF
REG’S
R/W
TYPE VALUES / REPRESENTATION / REMARKS
[d] 74
[h] 0x04A
40075
thermal
expansion
coefficient
2
R/W
uint32 0.000…9999.999*10
-6
/K
Representation: fixed at 9 decimals. Multiply returned value
by 10
-9
for original value
[d] 537
[h] 0x219
40538
T-normal (temp)
2
R/W
uint32 0.00…9999.99K