E126-P-EL
User Manual
FW_E126-P-EL_M_v0301-01_EN
Page 65
MODBUS-REGISTERS: TOTAL
PDU
ADDRESS
HOLDING
REGISTER
VARIABLE
NO. OF
REG’S
R/W
TYPE VALUES / REPRESENTATION / REMARKS
[d] 32
[h] 0x0020
40033
unit
1
R/W
uint16 0 = none
1 = L
2 = m3
3 = US GAL
4 = I GAL
5 = CF
6 = OILBBL
7 = kg
8 = ton
9 = lb
10 = US ton
[d] 33
[h] 0x0021
40034
decimals
1
R/W
uint16 0...3
[d] 34
[h] 0x0022
40035
K-factor
2
R/W
uint32 1...9999999
Representation 0.000010...9999999: decimals depending on
variable 37.
[d] 37
[h] 0x0025
40038
K-factor
decimals
1
R/W
uint16 0...6
[d] 47
[h] 0x002F
40048
multiply factor
1
R/W
uint16 0=x1
1=x10
2=x100
3=x1000
[d] 1052
[h] 0x041C
41053
clear total
password
1
R/W
uint16 000...999
Setting 000 disables the clear total password feature.
[d] 2146
[h] 0x0862
42147
contract hour
1
R/W
uint16 0...23 Hour
[d] 2208
[h] 0x08A0
42209
current day total
4
R
uint64 0.000 ... 99999999999
Representation: unit, decimals depending on variables 32,
33.
[d] 2216
[h] 0x08A8
42217
previous day
total
4
R
uint64 0.000 ... 99999999999
Representation: unit, decimals depending on variables 32,
33.
MODBUS-REGISTERS: FLOWRATE
PDU
ADDRESS
HOLDING
REGISTER
VARIABLE
NO. OF
REG’S
R/W
TYPE VALUES / REPRESENTATION / REMARKS
[d] 48
[h] 0x0030
40049
unit
1
R/W
uint16 0=none
1=mL
2=L
3=m
3
4=US GAL
5=I GAL
6=CF
7=OilBBL
8=nL
9=nm
3
10=SCF
11=mg
12=g
13=kg
14=ton
15=lb
16=US
17=rev
18=P
[d] 49
[h] 0x0031
40050
time unit
1
R/W
uint16 0=/sec
1=/min
2=/hour
3=/day
[d] 50
[h] 0x0032
40051
decimals
1
R/W
uint16 0…3
[d] 51
[h] 0x0033
40052
K-factor
2
R/W
uint32 1…9999999
Representation: 0.000010…9999999 depending on variable
54: decimals K-factor
[d] 54
[h] 0x0036
40055
K-factor
decimals
1
R/W
uint16 0…6
[d] 62
[h] 0x003E
40063
filter
1
R/W
uint16 1…99
[d] 71
[h] 0x0047
40072
period
1
R/W
uint16 0.1…99.9sec
MODBUS-REGISTERS: DISPLAY
PDU
ADDRESS
HOLDING
REGISTER
VARIABLE
NO. OF
REG’S
R/W
TYPE VALUES / REPRESENTATION / REMARKS
[d] 64
[h] 0x040
40065
display function
1
R/W
uint16 0=total
1=flowrate
2=acc.total
[d] 2147
[h] 0x863
42148
day totals
1
R/W
uint16 0=hidden
1=operate
2=off
[d] 42
[h] 0x02A
40043
toggle
1
R/W
uint16 0=disable
1=enable
[d] 80
[h] 0x050
40081
LCD update
time
1
R/W
uint16 0=fast
1=1sec
2=3sec
3=15sec
4=30sec
5=off
[d] 67
[h] 0x043
40068
backlight
brightness
1
R/W
uint16 0=off
1=20%
2=40%
3=60%
4=80%
5=100%