FLUID COMPONENTS INTL
APPENDIX B - GLOSSARY
Model ST98 Flow Meter
B - 4
Doc. No. 06EN003291 Rev. A
NUMBER
Menu Function.
S/N and CO No.
Only the factory has access to this area.
K. CONSTANTS SETUP
Menu Function.
Parameter Definitions
See the parameters in Menu Function D.
L. CALIBRATE OUTPUTS
Menu Function.
(U)p (P)down
Increases or decreases the DAC count.
(F)fast/slow
This controls the speed of the DAC counting.
M. MIN/MAX A/D LIMITS
Menu Function.
Max A/D
High end cut-off A/D number. Prevents false low flow readings.
Min A/D
Low end cut-off A/D number. Prevents false high flow readings.
N. SOFTWARE RESET
Menu Function.
Software Reset
Resets instrument without removing power.
O. SELECT SENSOR
HEATER CURRENT
Menu Function.
xxxLO
Choosing xxxLO sets software flags to show user if flow values are out of range.
xxxMD
Choosing xxxMD does not set any software flags.
P. RECONFIGURE
THE FC88 UNIT
Menu Function.
FC88 Reset
Re-configures the FC88 so it will function properly in conjunction with the ST98.
R. A/D CALIBRATE
RESISTANCE
Menu Function.
A/D Delta-R
The difference between the RTD resistances as used by the A/D converter.
A/D Ref-R
The reference RTD resistance as used by the A/D converter.
r = xxx
A/D Delta-R resistance
R = xxx
A/D Ref-R resistance.
S. AUTO SCALE
Menu Function.
Auto Scale ON
Always on.
T. NORMAL OPERATING
MODE
Menu Function.
U. DISPLAY TOTAL
FLOW TIME
Menu Function.
Time
Time in minutes since the last reset
Reset
Resets time to zero.
V. OUTPUT MODE
SELECT
Menu Function.
Output
Displays the selected instrument output (4-20 mA, 0-5 VDC, 0-10 VDC, 1-5 VDC).
W. TOTALIZER MODE
Menu Function.
Totalizer is
The Totalizer can be toggled on or off. If it is on, the results are displayed on the
second line of the normal operating mode window.
Reset Totalizer
The Totalizer can be reset to 0 with this command.
During Powerup
The Totalizer can be automatically reset each time power is applied to the
instrument.