7. Tighten tube seal clamp screw. Be sure to tighten firmly
to prevent “tubing walk.” Tubing walk can occur when
the tube seal is the wrong size or is not sufficiently tight
to keep the rollers from pulling the tubing through the
pump as it rotates.
8. Inspect pump cover gasket. Replace if damaged.
Reinstall pump cover gasket and cover.
Align the tubing marks
with the outside edge
of the Hose Seal.