The following is a list of abbreviations as they appear on the display and front panel of the unit.
ABBREVIATION
DESCRIPTION
FACTOR
SCALING FACTOR - For A and B Inputs. Each input has a separate 5 digit dividing scale factor.
DP F A
DECIMAL POINT FOR FACTOR A - Enter location of decimal point for scaling Factor A by pressing the
button under the digit where the decimal is desired.
DP F B
DECIMAL POINT FOR FACTOR B - Enter location of decimal point for scaling Factor B by pressing the
button under the digit where the decimal is desired.
COUNT
PORTION OF MENU FOR SETTING COUNTER VARIABLES AND “C” DISPLAY COMBINATIONS.
DP LOC
DECIMAL POINT LOCATION - Enter desired location of decimal by pushing the button under the digit
where the decimal is desired. Changing the decimal will change the decimal location in the A,B&C total
as well as the rate displays.
total
TOTAL - Allows the unit to display A ADD B or A SUBTRACT B (
a add b
or
a sub b
)
ratio
RATIO - Allows the unit to display A DIVIDED BY B X 100 to show percent of A to B or
A DIVIDED BY A PLUS B X 100 to show percent of A to A + B. (
A/B
or
A/A+B)
A SUB B
A SUBTRACT B - The number of scaled pulses from Input B are subtracted from the number of scaled
pulses from Input A. (-99999 TO 999999)
A ADD B
A ADD B - The number of scaled pulses from the A Input are added to the number of scaled pulses from the
B Input.
a / b
A DIVIDED BY B - The number of scaled pulses of input A are divided by the number of scaled pulses of
input B X 100 to give the percent ratio of A
÷
B.
a / a+b
A DIVIDED BY A PLUS B - The number of scaled pulses of input A divided by A plus the number of scaled
pulses of input B X 100 to give the percent ratio of A
÷
A + B.
RST al
RESET ALL - When the remote reset is negative edged triggered, A,B, & C total displays are reset to 0.
The reset button (rst) will only reset what is being displayed. If the C display is reset both A&B displays
are reset as well.
rsT 1
RESET ONE - Will give a chose between resetting the A or B total displays. If the A display is chosen and the
remote reset is negative edged triggered, only the A display is reset. If B display is chosen the same applies.
The C display will update accordingly. The reset button (rst) will only reset what is being displayed.
If the C display is reset both the A&B displays are reset as well.
HI CPS
HIGH COUNTS PER SECOND - This sets the unit to accept high count speeds (0-9.99KHz)
LO CPS
LOW COUNTS PER SECOND – This sets the unit for contact debounce filtering (0-40Hz)
RATE
PORTION OF MENU FOR SETTING RATE VARIABLES
RPS
RATE PER SECOND - The display will read in rate per second.
SCALE
SCALING - Allows unit to display rate per minute or rate per hour.
Z 60
DIVIDE K FACTOR BY 60 - This sets the unit for rate per minute; equal to 60 times rate per second.