EASYVAC VACUUM REGULATOR              MS003           February 2010 

 

Maintenance operations 

 

 

 

Replace the I-0 buttons group, complete with the 
relative sealing gaskets with caps. Spread a light 
layer of lubricant over all sealing elements of a 
new I-0 tap, using oxygen-compatible products 
(e.g. FOMBLIN YNX). 

Insert the green button on the left-hand side of the 
EASYVAC vacuum regulator, ensuring it is turned 
with the indication “I” towards the front. The rear 
slot of the button must fit into the relative coupling 
inside the body of the vacuum regulator itself. 

 

 

Insert the red button on the right-hand side of the 
EASYVAC vacuum regulator, ensuring it is turned 
with the indication “0” towards the front.  The rear 
slot of the button must fit into the relative coupling 
inside the body of the vacuum regulator itself. 

Keeping the green button pressed down, push 
down until the red button clicks fully into place. 
Move several times in both directions in order to 
lubricate the gasket housing completely and check 
that the I-0 switch moves correctly. Perform the 
functional checks required and indicated in the 
operating manual Mo_002 of the EASYVAC 
device before use. 

 WARNING: DO NOT USE OIL TO LUBRICATE THE PARTS.

 

 

2.3.2  CLEANING OF THE MEMBRANE AND THE BODY 

The vacuum pressure membrane and the one inside the body of the EASYVAC vacuum regulator should be 
cleaned periodically, to ensure the device keeps working properly. 
The cleaning methods are shown with the assistance of the following illustrations: 

Содержание EASYVAC

Страница 1: ...SERVICE MANUAL SERVICE MANUAL EASYVAC VACUUM REGULATOR PUBLICATION CODE MS003 EDITION no 6 DATE February 2010...

Страница 2: ...YVAC VACUUM REGULATOR MUST ALWAYS BE USED WITH CARE AND ONLY BY PERSONNEL WHO ARE AWARE OF THE CONSEQUENCES OF THE ONGOING THERAPY Disconnect the vacuum supply when the unit is not in use Cleaning and...

Страница 3: ...n of FLOW METER S p A FLOW METER S p A appliances are subjected to periodic revisions to maintain and improve the production standards functioning and increase their reliability For this reason the co...

Страница 4: ...SERVICE MANUAL 4 EASYVAC VACUUM REGULATOR MS003 February 2010 Table of illustrations Drawing no TITLE PAGE 1 1 Spares for EASYVAC vacuum regulator 5...

Страница 5: ...SERVICE MANUAL 5 EASYVAC VACUUM REGULATOR MS003 February 2010 Drawing no 1 1...

Страница 6: ...00 I O button switch complete with gaskets 3a 000010192 Vacuum gauge 1000 mbar for EASYVAC regulator complete with fixing support gaskets and protection cover 3b 000010193 Vacuum gauge 600 mbar for EA...

Страница 7: ...ry 6 months 2 3 3 REPLACEMENT OF THE LID WITH A MEMBRANE every 24 months 2 3 4 REPLACEMENT OF THE VACUUM GAUGE in the case of accuracy outside the tolerance levels or breakage 2 3 MAINTENANCE OPERATIO...

Страница 8: ...urned with the indication 0 towards the front The rear slot of the button must fit into the relative coupling inside the body of the vacuum regulator itself Keeping the green button pressed down push...

Страница 9: ...mpened with ethyl alcohol diluted to 60 in water Reattach the lid complete with the regulating knob and the membrane onto the body of the vacuum regulator tightening it manually for as long as resista...

Страница 10: ...he vacuum gauges they must be replaced as specified in the Table of illustrations on page 4 WARNING REPLACE THE VACUUM GAUGES USING ONLY ORIGINAL SPARE PARTS WITH THE SAME FULL SCALE VALUE AS THE ONES...

Страница 11: ...m gauge Carefully insert the vacuum gauge unit back into the body aligning the vacuum gauge holder block guides with those on the body of the EASYVAC vacuum regulator pushing it right in Be careful to...

Отзывы: