SERVICE
MANUAL
8
EASYVAC VACUUM REGULATOR MS003 February 2010
Maintenance operations
Replace the I-0 buttons group, complete with the
relative sealing gaskets with caps. Spread a light
layer of lubricant over all sealing elements of a
new I-0 tap, using oxygen-compatible products
(e.g. FOMBLIN YNX).
Insert the green button on the left-hand side of the
EASYVAC vacuum regulator, ensuring it is turned
with the indication “I” towards the front. The rear
slot of the button must fit into the relative coupling
inside the body of the vacuum regulator itself.
Insert the red button on the right-hand side of the
EASYVAC vacuum regulator, ensuring it is turned
with the indication “0” towards the front. The rear
slot of the button must fit into the relative coupling
inside the body of the vacuum regulator itself.
Keeping the green button pressed down, push
down until the red button clicks fully into place.
Move several times in both directions in order to
lubricate the gasket housing completely and check
that the I-0 switch moves correctly. Perform the
functional checks required and indicated in the
operating manual Mo_002 of the EASYVAC
device before use.
WARNING: DO NOT USE OIL TO LUBRICATE THE PARTS.
2.3.2 CLEANING OF THE MEMBRANE AND THE BODY
The vacuum pressure membrane and the one inside the body of the EASYVAC vacuum regulator should be
cleaned periodically, to ensure the device keeps working properly.
The cleaning methods are shown with the assistance of the following illustrations: