EASYVAC VACUUM REGULATOR              MS003           February 2010 

 

 

Maintenance operations 

 

 

 

Screw the block back onto the vacuum gauge, 
complete with the relative sealing gaskets, for as 
long as slight resistance between the two parts 
can be felt. Tighten for another ¼ turn, positioning 
the surfaces in alignment with the metal coupling 
of the vacuum gauge. 

Carefully insert the vacuum gauge unit back into the 
body, aligning the vacuum gauge holder block guides 
with those on the body of the EASYVAC vacuum 
regulator, pushing it right in. Be careful to ensure the 
sealing gasket remains inserted in the block, to avoid 
damage to it and a consequent loss of vacuum. 

 

 

Reposition the lid and fix it in place with the two 
screws by using a TORX screwdriver of suitable 
dimensions. Perform the functional checks 
required and indicated in the operating manual 
Mo_002 of the EASYVAC device before use. 

 

Содержание EASYVAC

Страница 1: ...SERVICE MANUAL SERVICE MANUAL EASYVAC VACUUM REGULATOR PUBLICATION CODE MS003 EDITION no 6 DATE February 2010...

Страница 2: ...YVAC VACUUM REGULATOR MUST ALWAYS BE USED WITH CARE AND ONLY BY PERSONNEL WHO ARE AWARE OF THE CONSEQUENCES OF THE ONGOING THERAPY Disconnect the vacuum supply when the unit is not in use Cleaning and...

Страница 3: ...n of FLOW METER S p A FLOW METER S p A appliances are subjected to periodic revisions to maintain and improve the production standards functioning and increase their reliability For this reason the co...

Страница 4: ...SERVICE MANUAL 4 EASYVAC VACUUM REGULATOR MS003 February 2010 Table of illustrations Drawing no TITLE PAGE 1 1 Spares for EASYVAC vacuum regulator 5...

Страница 5: ...SERVICE MANUAL 5 EASYVAC VACUUM REGULATOR MS003 February 2010 Drawing no 1 1...

Страница 6: ...00 I O button switch complete with gaskets 3a 000010192 Vacuum gauge 1000 mbar for EASYVAC regulator complete with fixing support gaskets and protection cover 3b 000010193 Vacuum gauge 600 mbar for EA...

Страница 7: ...ry 6 months 2 3 3 REPLACEMENT OF THE LID WITH A MEMBRANE every 24 months 2 3 4 REPLACEMENT OF THE VACUUM GAUGE in the case of accuracy outside the tolerance levels or breakage 2 3 MAINTENANCE OPERATIO...

Страница 8: ...urned with the indication 0 towards the front The rear slot of the button must fit into the relative coupling inside the body of the vacuum regulator itself Keeping the green button pressed down push...

Страница 9: ...mpened with ethyl alcohol diluted to 60 in water Reattach the lid complete with the regulating knob and the membrane onto the body of the vacuum regulator tightening it manually for as long as resista...

Страница 10: ...he vacuum gauges they must be replaced as specified in the Table of illustrations on page 4 WARNING REPLACE THE VACUUM GAUGES USING ONLY ORIGINAL SPARE PARTS WITH THE SAME FULL SCALE VALUE AS THE ONES...

Страница 11: ...m gauge Carefully insert the vacuum gauge unit back into the body aligning the vacuum gauge holder block guides with those on the body of the EASYVAC vacuum regulator pushing it right in Be careful to...

Отзывы: