FlowLine LP10 Series Скачать руководство пользователя страница 3

SAFETY PRECAUTIONS

Step Two

About this Manual:

PLEASE READ THE ENTIRE MANUAL PRIOR TO INSTALL-
ING OR USING THIS PRODUCT. This manual includes informa-
tion on all models of Intrusive RF Capacitance level switches from
Flowline, LP10 series. Please refer to the part number located on
the switch label to verify the exact model which you have pur-
chased.

User’s Responsibility for Safety:

Flowline manufactures a wide range of liquid level sensors and
technologies. While each of these sensors is designed to operate in
a wide variety of applications, it is the user’s responsibility to select
a sensor model that is appropriate for the application, install it
properly, perform tests of the installed system, and maintain all
components. The failure to do so could result in property damage
or serious injury.

Proper Installation and Handling:

Use a proper sealant with all installations. Never overtighten the
sensor within the fitting, beyond a maximum of 80 inch-pounds
torque. Always check for leaks prior to system start-up. Do not
install the LP10 series sensor within 6" of any metal.

Material Compatibility:

The LP10 series sensor is available in two different wetted materi-
als. Models LP10-1___ are made of Polypropylene (PP). Models
LP10-2___ are made of Perfluoroalkoxy, also known by the trade
name Teflon, (PFA). Make sure that the model you have selected
is compatible with the application liquid. To determine the chemi-
cal compatibility between the sensor and its application liquids,
refer to an industry reference such as the Compass Corrosion Guide
(available from Compass Publications, phone 619-589-9636).

Wiring and Electrical:

The supply voltage used to power the LP10 series sensor should
never exceed a maximum of 36 volts DC. Electrical wiring of the
sensor should be performed in accordance with all applicable
national, state, and local codes.

Flammable, Explosive and Hazardous Applications:

Only the LP10-___0 series switch is rated for use in hazardous
locations. Refer to the certificate of Compliance for all applicable
intrinsically safe ratings and entity parameters for the LP10-___0.
Refer to the National Electric Code (NEC) for all applicable
installation requirements in hazardous locations. DO NOT USE
THE LP10-___2, LP10-___3 or LP10-___5 GENERAL PUR-
POSE SWITCH IN HAZARDOUS LOCATIONS.

WARNING

Do not install the LP10 level switch on a metallic tank,
or within 6" of any metallic object. Metal will ad-
versely affect the dielectric sensitivity of the sensor.

The maximum current draw on the FET switch is 100 mA.

Flowline’s LP10 series sensors are not recommended
for use with electrically charged application liquids.
For the most reliable operation, the liquid being
measured will need to be electrically grounded.

INTRODUCTION

Step Three

About Intrusive RF Capacitance Technology:

Flowline’s LP10 series generates a 300 kHz pulse-wave radio frequency
signal from the capacitance electrode located in the sensing tip of each
sensor. When liquid comes into contact with the sensing tip, the
capacitance as measured by the sensor changes based on the dielectric
constant of the liquid. Flowline’s sophisticated electronics convert the
capacitance value into a simple two-wire, 4 or 20 mA, a three-wire FET
switch output which indicates whether the sensor is wet or dry or a 4-wire
NO or NC 1A relay output.  In the dry state, the sensor delivers a two
wire, 5 mA signal output. In the wet state, the sensor delivers a two wire,
19 mA signal output.

The sensor’s operation may vary based on the dielectric properties of
various application liquids. The LP10 series sensor is factory-calibrated
to be used with liquids with a dielectric value between 20 and 80. For
example, if the application liquid is acetal (with a dielectric constant of
about 3.6), the sensor (as factory calibrated) will not detect its presence
when it is wet. However, if the application liquid is glycol (with a
dielectric constant of about 35) the sensor will reliably detect its
presence when wet.

Warning:  Do not install the LP10 series sensor within 6 inches of any
metallic object. Metal will adversely affect the dielectric sensitivity of
the sensor.

Flowline’s LP10 series sensors are not recommended for use with
electrically charged application liquids. For most reliable operation, the
liquid being measured may need to be electrically grounded.

Acetone

21

Acetoaldehyde

22.2

Acetyl methyl hexyl

ketone

28

Alcohol

16 to 31

Ammonia

15 to 25

Acetic acid

4.1 to 6.2

Butyl chloride

9.6

Barium chloride

9 to 11

Benzene

2.3

Benzine

2.3

Barium nitrate

5.6

Bromine

3.1

Chlorobenzene

4.7 to 6

Chlorotoluene

4.7

Chloroform

4.5 to 5.0

Chlorine, liquid

2.0

Carbon tetrachloride

2.2

Cyan

2.6

Cyclohexane methanol 3.7
D.I. Water

20

Ethyl toluene

2.2

Ethyl alcohol

23

Ethylene glycol

37

Ethylene oxide

14

N-butyl formate

2.4

Nitrobenzene

26 to 35

Nitrotoluene

25

Naphthalene

2.3 to 2.5

Oils, vegetable 2.5 to 3.5
Oils, mineral

2.3 to 2.4

Oils, petroleum 1.8 to 2.2
Oleic acid

2.5

Propane, liquid 1.8 to 1.9
Potassium nitrate

5.0 to 5.9

Potassium chloride

5.0

Stearic acid

2.3

Toluene

2.4

Trichloroethylene

3.4

Trichloroacetic acid

4.5

Terephthalic acid

1.5 to 1.7

Thinner

3.7

Urea

3.5

Vinyl chloride

2.8 to 6

Vinyl alcohol

1.8 to 2.0

Water, 20

°

C

80

Water, 100

°

C

48

Ethylene dichloride

11 to 17

Ethylene chloride

10.5

Ethyl acetate

6.4

Ethyl salicylate

8.6

Ethyl stearate

2.9

Ethyl silicote

4.1

Formic acid

59

Ferric oleate

2.6

Freon

2.2

Glycerine

47

Glycol

30

Glycol nitrite

27

Gasoline

2 to 2.2

Hydrochloric acid

4.6

Isobutyric acid

2.7

Isobutyl methyl ketone 13
Jet fuel

1.7

Lead carbonate

18

Lead nitrate

38

Methyl salicylate

9.0

Methanol

33

Methyl alcohol

33 to 38

Margarine, liquid2.8 to 3.2
Methyl acetate

7.3

Table of Common Dielectric Constants

NOTE: Liquids with a dielectric constant less than 20 will not be
detected by an LP10 series sensor, as factory calibrated.