Simulation (SIM-T)
This mode enables the end user to simulate changes in level without having the level of liquid
move. The display will simulate level of liquid and will turn the relays ON and OFF according
to how they were configured. For example, if Relay 1 is set as a High Alarm with an ON setting
for 100.0 inches and a Hysteresis of 2.0 inches, then when the display goes above 100.0, Relay 1
will energize and when the level falls below 98.0, Relay 1 will de-energize. This function can be
used to test the wiring and to verify if the configuration is set accordingly.
Please note: Simulation mode will energize and de-energize all configured relays. Any devices
wired to the relays will become active during simulation mode.
Follow the instructions below to begin the Simulation mode (SIM-T)
1.
From the main MENU level, press SETUP when HELP appears.
2.
When SIM-T appears, press the SELECT key.
3.
The display will start at 0.0. Press the UP button to begin increasing the level of liquid.
Use the DOWN button to simulate the lowering of liquid in the tank.
4.
To increase speed, hold the SELECT button while hold the UP or DOWN keys.
5.
To exit, press and hold the SELECT key.
6.
Press SELECT when appears EXIT to return to the main program level.
Test Parameters
This mode runs diagnostic tests that confirm operation of EchoSwitch II.
This mode should
only be used when supervised by a Flowline representative.