TECHNICAL DATA
68
Flow Meter |
FC
01- CA
9 .5 .2 Temperature measurement:
Measuring range:
- 40 … +130 °C
Accuracy:
±1 % MB ***
9 .5 .3 Electronic control unit FC01-CA
Temperature drift:
0.1 %/K/MBE * (CSP-..)
0.05 %/K/MBE * (CSF-, CST-…)
Warm up period until full
accuracy is reached:
15 min.
* MBE - of final value
** MW - measured value
*** MB - measuring range
9 .6 Maintenance
The sensor is maintenance free for fluids that do not adhere to the sensor tips. If impurities or particles
are present in the fluid and adhere to the sensor tips, this can cause incorrect measured values. In
this case, the sensor tips must be cleaned at suitable intervals. When cleaning, make sure that the
sensor tips are not damaged.