INSTALLATION
27
FC
01- CA
| Flow Meter
XAO - Analogue outputs
Connection by 8 pole connector; max. 1.5 mm
2
; LiYCY 2 x 0.25 mm
2
cable recommended
for each analogue output
Pin selection for analogue outputs
(option: V1, V2, C1)
Pin No.
Signal name
Function
1
nc
none
2
ANAO1
analogue output 1 - flow
3
ANA1GND
reference potential for analogue output 1
4
SGNDA1
shield ground for analogue output 1 (ungrounded) *
5
SGNDA2
shield ground for analogue output 2 (ungrounded) *
6
ANAO2
analogue output 2 - temperature
7
ANA2GND
reference potential for analogue output 2
8
nc
none
* Apply shield on one side only .
XAH - Limit switch signal outputs - relay outputs - single pole double throw
Connection by 8 pole connector; max. 1.5 mm
2
; LiYCY 3 x 0.38 mm
2
cable recommended
for each signal output
Pin No.
Signal name
Function
1
SGNDL1
shield ground 1
2
LIM1
non-inverted signal output 1 (N.O.)
3
LIM1COM
common 1
4
/LIM1
inverted signal output 1 (N.C.)
5
SGNDL2
shield ground 2
6
LIM2
non-inverted signal output 2 (N.O.)
7
LIM2COM
common 2
8
/LIM2
inverted signal output 2 (N.C.)