![Flomotion Systems LT260 Скачать руководство пользователя страница 13](http://html.mh-extra.com/html/flomotion-systems/lt260/lt260_operation-and-maintenance-manual_2306734013.webp)
13
[P05] Dead Zone
Sets the distance from the face of the transducer to the closest readable level.
Setting
Default
Minimum 01.00 - 26.00ft (0.30 - 7.92m)
01.00ft
[P06] 4mA Set Point
[Default: 00.00ft]
Sets the 4mA value for the measurement type selected to P01. 4mA typically is Empty (0% of
Span)
Setting
Default
Minimum 00.00 - 26.00ft (0 - 7.92m)
00.00ft
[P07] 20mA Set Point
[Default: 26.00ft]
Sets the 20mA value for the measurement type selected in P01. 20mA typically is Full (100% of
Span).
Setting
Default
Minimum 00.00 - 26.00ft (0 - 7.92m)
26.00ft
[P08] mA Fail-Safe Value
[Default: 1 (3.8mA)]
Upon Loss of Echo, the mA output can be set to indicate a fault condition.
Option
1
3.8 mA: Fault condition (LOE) indicated by 3.8mA
2
Hold: Last know measured value
3
22mA: Fault condition (LOE) indicated by 22mA
[P09] mA Fail-Safe Time
Sets the time the Fail-Safe value is held.
Setting
Default
Minimum 20 - 900 sec
300 sec
Содержание LT260
Страница 1: ...LT260 Ultrasonic Level Transmitter Operation and Maintenance Manual ...
Страница 2: ...2 ...
Страница 17: ...17 ...