INSTALLATION

EARTH GROUND

When making installations, the 
magnetic flow meter grounding rules 
have to be observed. The sensors 
of the QSE flow meter are sensitive 
to any electrical noise that is always 
present in piping systems. The fluid 
in piping systems usually carries 
significant levels of static electricity 
and should be grounded for the QSE 
meter to be at its best performance.

As part of the installation, it is 
important to understand the 
importance of having the QSE meter 
connected to “earth ground”. Earth 
grounding helps ensure electronic 
component stability and reliability 
by using the earth to absorb any 
static charge buildup or spurious 
signal noise that can affect the meter 
electronics. Earth ground in this 
installation is NOT the ground wire in 
the customer isolated power supply.

Process Fluid Grounding

Establishing a process fluid ground 
is one of the most important 
installation details. Proper process 
fluid grounding ensures that the 
sensor and fluid are at the same 
electrical potential so that only the 
induced flow signal is measured. The 
process fluid grounding provides 
a stable base potential reference 
for measuring the induced flow 
signal (the differential between the 
base potential reference and the 
fluid flow potential created by the 
sensor magnetic field). Process fluid 
grounding should be established 
on both sides (upstream and 
downstream) of the sensor elements.

By connecting the QSE sensor, 
the fluid, and the reference used 
by the internal electronics to a 
stable, noise free reference point 
(earth ground), the user is ensured 
of getting the best performance 
from their magnetic flowmeter.

Each QSE meter has a ground 
lug with a removable ring terminal 
located in the circular wall adjacent 
to the outlet end of the meter. This 
ground lug is connected internally 
to the electronics and sensors and 
is the meter connection to earth 
ground (the base potential reference).

Grounding Summary:

1.  Ground sensor to meter 

ground lug. (Done at factory.)

2.  Ground electronics to meter 

ground lug. (Done at factory.)

3.  Ground process fluid to 

meter ground lug. (Customer 
responsibility.) Various process 
fluid grounding examples are 
shown on the following pages 
(

see Figures 1,2, 3, and 4

) to 

assist you in completing this 
item. The examples cover various 
meter installation situations.

4.  Ground meter ground lug to earth 

ground.  (Customer responsibility)

All of the examples on the following 
pages show 14-16 AWG ground 
wire (customer supplied) being 
used. A ground wire should be 
crimped to the ring terminal on the 
meter ground lug and the other 
end of the ground wire connected 
to earth ground (item 4 above).

Содержание QSE Series

Страница 1: ...oused within either of the two cover plates A display Q09 is also available mounted to the display mount cover plate All meters are equipped with galvanically isolated pulse out electronics QSB as the default standard regardless of style of cover plate This manual contains overall information related only to the meter This meter is externally powered and all external wiring connects to the electro...

Страница 2: ...ic noise If the QSE series meters are used in a manner not specified by the manufacturer the protection provided by the equipment may be impaired PRINCIPLE OF OPERATION Faraday s Law of Electromagnetic Induction is the operating principle on which the QSE series meters are based Faraday s Law paraphrased states that a voltage will be induced in a conductor when it passes through a magnetic field a...

Страница 3: ...e potential reference and the fluid flow potential created by the sensor magnetic field Process fluid grounding should be established on both sides upstream and downstream of the sensor elements By connecting the QSE sensor the fluid and the reference used by the internal electronics to a stable noise free reference point earth ground the user is ensured of getting the best performance from their ...

Страница 4: ...2 PLACES CONNECT WIRE TO RING TERMINAL TIGHTEN RING TERMINAL WITH SCREW QSE GROUNDING LUG QSE GROUNDING LUG A CONNECT WIRE TO RING TERMINAL TIGHTEN RING TERMINAL WITH SCREW 3 PLACES EARTH GROUND SIDE VIEW WITH CUT AWAY DETAIL A CUT AWAY VIEW TYPICAL 2 PLACES SCREW HAND TIGHT RING TERMINAL GROUNDING PROBE APPLY THREAD TAPE OR PIPE SEALANT PRIOR TO INSTALL 1 8 27 NPT TO BE TAPPED INTO PORT CAVITY Ø ...

Страница 5: ...SEGMENTS PLASTIC PIPE FLANGE TYP METAL PIPE FLANGE TYP GROUNDING SCREW TYP QSE GROUNDING LUG QSE GROUNDING LUG QSE GROUNDING ANSI FLANGE NON CONDUCTIVE PLASTIC PIPE USING COMMERCIALLY AVAILABLE GROUNDING RINGS QSE GROUNDING ANSI FLANGE CONDUCTIVE METAL PIPE USING INTEGRATED CONTINUITY GROUND WIRE METER FLANGE TYP GASKET TYP GASKET TYP GROUNDING RING TYP COMMERCIALLY AVAILABLE EARTH GROUND CONNECT ...

Страница 6: ... 20 Electrode Locations 5 x DIA 3 x DIA FLOW DIRECTION Figure 5 Meter shown tilted 20 degrees in horizontal position Plan to install meter with minimum straight pipe lengths at inlet and outlet ends The straight run lengths noted below represent the minimum requirements for accurate flow measurement see Figure 5 For optimum performance provide as much additional straight run as possible Upstream f...

Страница 7: ...installing the meter install an O ring in the O ring gland on each housing end then fit a backup ring over each O ring see Figure 6 The BSPP meter is now ready for installation Hand tighten the meter at the housing ends DO NOT use a wrench or similar tool to tighten as this can damage the housing 1 Install O ring 2 Install backup ring Installed Flush to center O ring with housing cut away view Fig...

Страница 8: ...our meter may have two 1 8 in NPT pipe plugs installed on top of the flow tube near each end They have been installed and sealed at the factory and are for use only with optional temperature sensing probes for energy use calculations in specialized applications RECOMMENDED INSTALLATION WARNING Compatibility of this product s material and the process fluid and or environment should be considered pr...

Страница 9: ...d strain reliefs will accommodate a cable diameter of 0 11 0 26 inches 2 79 6 6mm and provide an environmental seal around the cable when the dome nut is tightened NPT Adapters The 1 2 NPT adapter fittings are used for attaching flex conduit to the meter for those applications that require cable runs to be enclosed in conduit Cable to be provided by customer to accommodate job requirements Cable i...

Страница 10: ...ll of liquid Do not allow meter to discharge to atmosphere Conductivity of fluid too low Consult liquid properties Operating outside meter limits Increase decrease flow rate to proper meter specification Review temperature limits Electrical noise Install meter away from devices that emit EMF Ground loop Meter may need to be earth grounded Electrodes not clean Clean electrode tips Interface device ...

Страница 11: ...nge Steel WETTED COMPONENTS Housing NORYL GFN3 PPE PS Electrodes 316L Stainless Steel Temperature Probes 316 Series Stainless Steel Pipe Plugs 300 Series Stainless Steel O Rings EPDM Ethylene Propylene Diene Monomer MAXIMUM WORKING PRESSURE De rate maximum working pressure 0 682 PSIG per each degree F above 70 F NPT 150 PSIG 70 F 10 3 BAR 21 C BSPP 150 PSIG 70 F 10 3 BAR 21 C ANSI FLANGE 150 PSIG ...

Страница 12: ...m 400 3 inch 5 to 300 gpm 25 to 15 fps 0 5 of Reading 0 003 fps 060 gpm 121 4 inch 10 to 600 gpm 25 to 15 fps 0 5 of Reading 0 003 fps 120 gpm 68 1 FLUID OPERATING TEMPERATURE 32 F to 210 F NPT BSPP 32 F to 180 F Polymer and Steel Flanges Do not allow fluid to freeze inside meter STORAGE TEMPERATURE 40 F to 228 F AMBIENT AIR OPERATING TEMPERATURE 0 F to 140 F ALL METERS There is a correlation betw...

Страница 13: ... 0 mm s 114 L min 105 7 3 inch 19 to 1140 L min 076 to 4 57 m s 0 5 of Reading 1 0 mm s 227 L min 30 4 inch 38 to 2270 L min 076 to 4 57 m s 0 5 of Reading 1 0 mm s 454 L min 18 FLUID OPERATING TEMPERATURE 0 C to 98 C NPT BSPP 0 C to 82 C Polymer and Steel Flanges Do not allow fluid to freeze inside meter STORAGE TEMPERATURE 40 C to 108 C AMBIENT AIR OPERATING TEMPERATURE 18 C to 60 C ALL METERS T...

Страница 14: ...0 2 4 in 7 4 13 4 10 8 Weight with display For plain cover plates subtract 0 09 kg PRODUCT WEIGHT lb NPT BSPP ANSI Polymer Flange ANSI Steel Flange DIN Steel Flange 1 2 in 2 5 3 4 in 2 6 1 in 2 7 1 1 2 in 3 9 2 in 4 3 3 in 14 24 4 22 5 4 in 16 3 29 5 23 7 Weight with display For plain cover plates subtract 0 2 lb U S PRODUCT WEIGHT METRIC PRODUCT WEIGHT ...

Страница 15: ...ECTRONICS A B E D C MILLIMETER DIMENSIONS SHOWN IN BRACKETS 6 31 160 2 ALL METERS DIMENSIONS QSE METER DIMENSIONS NPT BSPP ANSI FLANGE DIN FLANGE listed in inches millimeters listed in brackets METER SIZE FITTING A B C D E 1 2 in NPT BSPP 5 20 132 1 5 83 148 1 10 50 266 7 5 13 130 9 1 83 46 5 3 4 in NPT BSPP 5 20 132 1 5 83 148 1 10 75 w273 1 5 13 130 9 1 83 46 5 1 in NPT BSPP 5 20 132 1 5 83 148 ...

Страница 16: ...16 TEMPERATURES ...

Страница 17: ...EMPERATURE F CONDUCTIVITY µs cm Coca Cola Syrup 68 600 Coffee Extract 183 2 5000 Corn Syrup 89 6 16 Ethylene Glycol Pure 68 1 07 Gin 90 Proof 77 10 ISO propyl Alcohol 77 3 5 Molasses 50 300 RC Cola Syrup 77 600 Sugar Solution Dilute 86 585 Urea 77 5000 Vodka 100 Proof 77 4 Water Distilled 0 04 Water NYC 77 72 NOTE This table of fluid conductivity is for reference only to show the relative conducti...

Страница 18: ...18 BSPP METER FITTING END L N1 N2 N3 W T B A1 A2 U P1 P2 P3 S1 S2 S3 S4 S5 S6 Polymer flanges shown V1 V2 V3 V4 V5 R C D E F G H J K M1 M2 M3 X ...

Страница 19: ...play 4 in Includes Computer display w decal seal 10 pin connector 1 K 145500 14 Kit QSB w display cover plate Includes Display cover plate QSB electronics with ribbon cable 6 screws cover plate seal 1 L 145500 15 Kit QSB w plain cover plate Includes Plain cover plate w decal QSB electronics with ribbon cable 6 screws cover plate seal 1 M1 145502 01 Kit QSI1 w display cover plate Includes Display c...

Страница 20: ... flange steel DN100 PN10 Includes 2 4 in steel flanges 4 4 in half ring keys 1 T 145500 12 Kit strain relief cover plate Includes 4 PG7 thread strain reliefs 11 26 cable diameter range 1 U 145500 13 Kit strain relief temperature sensor Includes 2 PG7 thread strain reliefs reduced fit 08 20 cable diameter range 1 V1 145500 20 Kit O ring backup ring 1 2 in BSPP Includes 2 O rings 2 backup rings 1 V2...

Страница 21: ...rk unless they are completely free of liquid residue The Waste Electrical and Electronic Equipment WEEE directive 2002 96 EC was approved by the European Parliament and the Council of the European Union in 2003 This symbol indicates that this product contains electrical and electronic equipment that may include bat teries printed circuit boards liquid crystal displays or other components that may ...

Страница 22: ...22 NOTES ...

Страница 23: ...23 NOTES ...

Страница 24: ...ed outside the warrantor s duly appointed representative B the product has been subjected to neglect misuse abuse or damage or has been in stalled or operated other than in accordance with the manufacturer s operating instructions To make a claim against this warranty contact the GPI Customer Service Department at 316 686 7361 or 888 996 3837 Or by mail at Great Plains Industries Inc 5252 E 36th S...

Отзывы: