INSTALLATION GUIDELINES
1.
GENERAL
The flow profile must be uniform at the point where the Insertion Meter is to be installed, otherwise
inaccurate and unstable readings will result.
British Standard: BS 1042 gives a full insight into flow conditioning for inferential flow devices
The general rule is to have a minimum of 10 diameters of straight pipe runs before the flowmeter
(upstream) and 5 diameters after (downstream). These straight runs must not contain any other items
such as Valves, Bends, Tees, Probes, or Reducers etc.
Any Valves immediately outside the straight runs require doubling of the straight pipe run lengths.
2.
COMMON CAUSES OF NON-UNIFORM FLOW