![Flo-tech mc4000 Скачать руководство пользователя страница 9](http://html1.mh-extra.com/html/flo-tech/mc4000/mc4000_user-manual_2306497009.webp)
Operation
Programming Menu
Press
MENU
to open the programming menu . Continue to press
MENU
to scroll through the user settings and measurement
parameters on the display . Press
UP
or
DOWN
to scroll through the options available for each parameter . Press
OK
to store
any new settings, the display will read
DATA STORED
. Press
ESC
to scroll backwards through the menu choices . Press
ESC
three
times to return to measuring mode .
Datalogger
The datalogger parameter allows you to start, stop or delete recorded measurement sessions .
DATALOGGER
STORE OFF
This command stops recording the datalogging information to memory . Press
OK
to stop
an active datalogging session . The screen will display
DATA STORED
.
DATALOGGER
STORE ON
This command starts recording data at a selected interval rate, see
below .
Each record automatically adds the date and time from an internal real time clock . To start a
new datalogging session press
OK
. The screen will display
DATA STORED
.
DATALOGGER
STORE DELETE
This command deletes all stored data . To delete data press
OK
, the unit will display
ERASE
DATA?
Press
OK
again to confirm the request . The screen will display
DATA ERASED
.
The internal memory can save all displayed measurements and permits 224 individual records at a total capacity of 2 .5 MB .
To start a new datalogging session, press
OK
at the
STORE ON
prompt . After starting a new datalogging session, press
ESC
to
switch the display to measuring mode . In measuring mode, an
M
icon indicates an active datalogging session, see
Stored datalogging sessions can be uploaded to a computer, using the USB data port, for further processing .
P1
bar
74.32
T
°C
86.5
Q
l/min
38.4
P2
bar
76.12
M
Logging Active
Indication
Figure 5: Active datalogging session display
Save Interval
The save interval is the time between two consecutive recording cycles set in fixed increments as shown below .
SAVE INTERVAL
1 s
Selection
Function
1s, 2s, 5s, 15s, 30s,
60s, 120s, 300s, 600s,
1200s, 1800s, 2700s,
3600s, 7200s
This command sets the time, in seconds, between two recording
cycles . Press
UP
or
DOWN
to scroll through each interval, when
the correct interval is displayed, press
OK
to select the interval . The
screen will display
DATA STORED
.
Page 9
February 2015
TST-UM-00017-EN-03