MC4CON Software
MC4000 Setup Tab
Under the
Menu MC4000
tab, all instrument parameters can be set and then uploaded to the MC4000 . The MC4000 also
transfers the process parameters and saved data currently displayed on this screen to the PC .
Menu MC4000 Setup
Menu MC4000 Setup
Exit
File
Com
Select
Info
Language
Flow-RPM
File
New device
File read
File save
Menu MC4000 read
Menu MC4000 store
Menu MC4000 write
RPM
Dscale RPM
Scale RPM
Sel.Flow-RPM
Select RPM
Select Flow
Select off
New-Lin
Dscale RPM
Scale RPM
Units
US
EU
Language
Spanish
English
Backlight
Light on
Light off
Sel.Flow type
160 gpm
80 gpm
40 gpm
7 gpm
New-Lin
New-Tab
Scale P1
Scale P2
Pressure
New-Tab
Tab F1 INP
Tab F2 INP
Tab F3 INP
Tab Q1 OUTP
Tab Q 2 OUTP
Tab Q3 OUTP
Save Interval
1 s
2 s
5 s
15 s
30 s
60 s
120 s
300 s
600 s
1200 s
1800 s
2700 s
3600 s
7200 s
Threshold
SIGNAL > 10%
SIGNAL < 10%
SIGNAL > 20%
SIGNAL < 20%
SIGNAL > 30%
SIGNAL < 30%
SIGNAL > 40%
SIGNAL < 40%
SIGNAL > 50%
SIGNAL < 50%
SIGNAL > 60%
SIGNAL < 60%
SIGNAL > 70%
SIGNAL < 70%
SIGNAL > 80%
SIGNAL < 80%
SIGNAL > 90%
SIGNAL < 90%
Device
Name
Datalogger
Menu MC4000
Measurement
Figure 12: Main MC4000 setup screen
Sel. Flow-RPM
Select flow and rpm settings.
Select off
Flow measurement display deactivated (display shows ----)
Select Flow
Flow measurement is active
Select RPM
RPM measurement is active
Sel. Flow type
Select the flow rate for the currently connected flow sensor.
7 gpm (25 lpm)
Default parameters for 7 gpm (25 lpm) flow sensor
40 gpm (150 lpm)
Default parameters for 40 gpm (155 lpm) flow sensor
80 gpm (300 lpm)
Default parameters for 80 gpm (300 lpm) flow sensor
160 gpm (600 lpm)
Default parameters for 160 gpm (600 lpm) flow sensor
New-Lin
Two point calibration for linear flow meter.
Scale Flow
Scale the
multiplication constant .
See
"Two Point Flow Sensor Calibration Using the New Lin Function" on page 14.
Dscale Flow
Scale the division constant .
Pressure
Pressure parameter settings
Scale P1
Full scale for the first pressure sensor (P1)
Scale P2
Full scale for the optional pressure sensor (P2)
Save Interval
Sampling and recording rate during data collection .
Threshold
The threshold signal level for the P1 (Transient) . Defines the point at which transient
recording starts
Page 19
February 2015
TST-UM-00017-EN-03