Name
Value
Protection
Peak voltage 500 V
Clamping voltage 150 V
Rated voltage 24 V
Electrostatic discharge (ESD) contact 8 kV
ESD air 15 kV
Connector type
Coaxial BNC jack
8.10
IRIG-B
The FLIR X6570sc features an IRIG-B input BNC connector (colored blue) on the back
panel of the camera (8 in Figure 5.3
Camera back panel description.
The value of the signal is digitalized during the integration of the infrared image and em-
bedded within it. It is then recorded by ResearchIR Max and stored in the sequence file
image headers.
The supported IRIG-B formats are IRIG-B12x. The signal should be 3:1, 3 Vpp maximum
at 50 Ω or 6 Vpp for high impedance input.
#T810204; r. AA/43079/43091; en-US