FLIR Extech Instruments Oyster-10 Скачать руководство пользователя страница 8

Oyster-10, V1.9 11/09 

8

Troubleshooting an Electrode Problem 

Symptom

 

Cause

 

Recommended Solution

 

Long response time or 
reading drift 

Clogged Junction 

Soak in 4.07 M Potassium Chloride 
(KCl) @ 60

o

C for 30 minutes. Replace 

electrode if necessary. 

Strong Alkaline 
measurement 

Soak in 0.1 M HCL overnight 

Deteriorated Gel layer of 
pH sensing membrane 

Replace electrode 

Protein coating on 
electrode surface 

Soak in 1 gm Pepsin dissolved in 
100ml of 0.1 m HCL for 30 mins. or as 
needed 

Oil, paint, dyes, suspended 
solids on sensor 

Rinse electrode alternately with 
materials solvent (briefly) then a 
buffer 7.00 

Organic solvents are 
coating sensor 

Organic mole fraction has to be less 
than 50% to assure reasonable 
readings. Limit time of measurement. 
Keep probe in the 7.0 buffer between 
readings. 

Dehydrated membrane 

Read solution for dry bulb below 

Dry Bulb 

Long term storage without 
wetting  

Soak electrode tip in wetting cap filled 
with 1ml 7.00 buffer for 24 to 48 hours

Static Charge 

Wiping electrodes 

Rinse electrode in 7.0 buffer and blot. 
Do no wipe electrode. 

Same readings in 
different buffers and 
samples 

Cracked or broken bulb 

Replace electrode. Use bulb guard. 
Avoid plunging electrode to bottom of 
container and spinning bars. A wetting 
cap will protect bulb between 
measurements. 

Erratic LCD display 

Samples have low ionic 
strength (lacks salt); e.g. 
distilled, de-ionized, boiled, 
lake water (high pressure) 

For each 50 ml of sample add 1 drop 
(50uL) of SAT.KCL No alteration in 
pH will occur by inert KCL. 

Содержание Extech Instruments Oyster-10

Страница 1: ...adings can be conveniently stored for later recall The Oyster 10 is battery operated portable and includes a hinged display that can be adjusted to any viewing angle In addition the neck strap offers...

Страница 2: ...ter Auto Power OFF After 10 minutes of inactivity Low Battery Indication Battery icon appears on LCD Operating Temperature 41 F to 104 F 5 C to 40 C Storage Temperature 4 o F to 140 o F 20 o C to 60 o...

Страница 3: ...e below The meter automatically recognizes all of the buffers listed in the table below and calibrates itself to the buffer s pH value 1 Connect the pH electrode to the meter s BNC connector input 2 O...

Страница 4: ...ater 8 Place approx 1 ml of pH 4 solution or tap water in the wetting cap before replacing it mV ORP Measurements 1 Note that calibration is not required for mV measurements 2 Use the pH mV button to...

Страница 5: ...ear 5 To clear the MAX MIN memory cycle the meter power External Temperature Offset Adjustment The Offset Adjustment is used to fine tune the temperature reading of the external temperature probe This...

Страница 6: ...o a sample buffer or rinse solution shake the electrode with a snap motion to remove residual drops of solution 6 Use part of the sample buffer to be tested next as a rinse solution whenever possible...

Страница 7: ...d replace the battery cover Troubleshooting the Meter Meters are the least likely cause of pH measurement problems To rule out the meter check the following 1 Check the 9V battery and replace it if ne...

Страница 8: ...fraction has to be less than 50 to assure reasonable readings Limit time of measurement Keep probe in the 7 0 buffer between readings Dehydrated membrane Read solution for dry bulb below Dry Bulb Lon...

Страница 9: ...rance and proper packaging to prevent damage in transit This warranty does not apply to defects resulting from action of the user such as misuse improper wiring operation outside of specification impr...

Отзывы: