FLIR Extech Instruments Oyster-10 Скачать руководство пользователя страница 6

Oyster-10, V1.9 11/09 

6

pH Electrode Model 60120B Operation 

The Following information applies to the Epoxy Body, Sealed Reference Combination pH 
Electrode Model 60120B. 

Electrode Considerations 

1.  The 60120B pH electrode is recessed in the polymer body allowing it to rest against 

the bottom of a beaker without damaging the glass bulb. This eliminates the need for 
electrode holders and allows the probe to be used as a stirring rod. This sealed 
reference design eliminates the need for adding filling solutions (minimizing "dry-out") 
and can be used in up to 100 psig systems without external pressurization. 

2.  The Electrode should remain in its bottle or cap until used. 
3.  If bubbles are seen in the bulb area, hold the electrode by its cap and shake 

downwards until bubbles disappear. 

4.  To improve speed of response, vigorously stir the electrode in the sample, buffer, or 

rinse solution. 

5.  After exposure to a sample, buffer, or rinse solution, shake the electrode with a snap 

motion to remove residual drops of solution. 

6.  Use part of the sample/buffer to be tested next as a rinse solution whenever possible. 
7.  When calibrating, use a buffer as close as possible to the expected pH of the sample. 
8.  Keep buffers and samples at the same temperature to avoid temperature effects. 
9.  pH readings stabilize at various rates; allow time for readings to stabilize. 
10. Electrodes deteriorate over time. If accuracy falls to 10% (slope), the electrode should 

be cleaned. If no improvement is observed, replace the electrode. 

Electrode Calibration and Cleaning 

Refer to the earlier section ‘Calibration and Operation’ for calibration details. Refer to the 
section below ‘Maintenance and Troubleshooting’ for details on electrode cleaning. 

Electrode Storage 

The electrode can be stored in the soaker bottle or wetting cap. For the storage bottle, 
slide the cover onto the electrode, replace the O-ring, and insert electrode into the bottle 
firmly tightening the cap. Ensure that the bottle or cap is filled with a pH of 4 buffer solution. 

Содержание Extech Instruments Oyster-10

Страница 1: ...adings can be conveniently stored for later recall The Oyster 10 is battery operated portable and includes a hinged display that can be adjusted to any viewing angle In addition the neck strap offers...

Страница 2: ...ter Auto Power OFF After 10 minutes of inactivity Low Battery Indication Battery icon appears on LCD Operating Temperature 41 F to 104 F 5 C to 40 C Storage Temperature 4 o F to 140 o F 20 o C to 60 o...

Страница 3: ...e below The meter automatically recognizes all of the buffers listed in the table below and calibrates itself to the buffer s pH value 1 Connect the pH electrode to the meter s BNC connector input 2 O...

Страница 4: ...ater 8 Place approx 1 ml of pH 4 solution or tap water in the wetting cap before replacing it mV ORP Measurements 1 Note that calibration is not required for mV measurements 2 Use the pH mV button to...

Страница 5: ...ear 5 To clear the MAX MIN memory cycle the meter power External Temperature Offset Adjustment The Offset Adjustment is used to fine tune the temperature reading of the external temperature probe This...

Страница 6: ...o a sample buffer or rinse solution shake the electrode with a snap motion to remove residual drops of solution 6 Use part of the sample buffer to be tested next as a rinse solution whenever possible...

Страница 7: ...d replace the battery cover Troubleshooting the Meter Meters are the least likely cause of pH measurement problems To rule out the meter check the following 1 Check the 9V battery and replace it if ne...

Страница 8: ...fraction has to be less than 50 to assure reasonable readings Limit time of measurement Keep probe in the 7 0 buffer between readings Dehydrated membrane Read solution for dry bulb below Dry Bulb Lon...

Страница 9: ...rance and proper packaging to prevent damage in transit This warranty does not apply to defects resulting from action of the user such as misuse improper wiring operation outside of specification impr...

Отзывы: