6.1
Procedure
Follow this procedure:
1. Put a battery into the battery compartment.
2. Connect the USB battery charger to the USB connector at the top of the camera.
3. Charge the battery for 2 hours before starting the camera for the first time.
4. Insert a memory card into the card slot at the top of the camera.
Note
Empty or use a memory card that has not previously been used in another type
of camera. The cameras may organize files differently on the memory card. There is
therefore a risk of losing data if the same memory card is used in different types of
cameras.
5. Push the on/off button
to turn on the camera.
6. Aim the camera toward the object of interest.
7. Adjust the infrared camera focus by rotating the focus ring.
Note
It is very important to adjust the focus correctly. Incorrect focus adjustment af-
fects how the image modes work. It also affects the temperature measurement.
8. Pull the trigger to save an image.
9. Download and install FLIR Tools/Tools+ (freeware) or FLIR Report Studio (licensed
software) on your computer.
10. Start FLIR Tools/Tools+ or FLIR Report Studio.
11. Connect the camera to the computer using the USB cable.
12. Import the images into FLIR Tools/Tools+ or FLIR Report Studio and create an inspec-
tion report.
13. Send the inspection report to your client.
6.2
To keep in mind
• Adjust the focus first. When the camera is out of focus, the measurement will be wrong.
• By default, most cameras adapt the scale automatically. Use this mode first, but do not
hesitate to set the scale manually.
• A thermal camera has a resolution limit. This depends on the size of the detector, the
lens, and the distance to the target. Use the center of the spot tool as a guide to the
minimum possible object size, and get closer if necessary. Make sure to stay away from
dangerous areas and live electrical components.
• Be careful when holding the camera perpendicular to the target. Be observant of reflec-
tions, especially at low emissivities—you, the camera, or the surroundings may become
the main source of reflection.
• Select a zone of high emissivity, e.g., an area with a matte surface, to perform a
measurement.
• Blank objects, i.e., those with low emissivities, may appear warm or cold in the camera,
because they mainly reflect the environment.
• Avoid direct sunlight on the details that you are inspecting.
• Various types of faults, e.g., those in a building’s construction, may result in the same
type of thermal pattern.
• Correctly analyzing an infrared image requires professional knowledge about the
application.
#T810190; r. AJ/46209/46209; en-US
12
Содержание E75
Страница 2: ......
Страница 3: ...User s manual FLIR Exx series T810190 r AJ 46209 46209 en US iii ...
Страница 4: ......
Страница 52: ...Handling the camera 11 T810190 r AJ 46209 46209 en US 40 ...
Страница 55: ...Handling the camera 11 11 15 Lanyard strap 11 15 1 General T810190 r AJ 46209 46209 en US 43 ...
Страница 172: ...Technical data 27 T199043 FLIR ResearchIR Max 4 Upgrade printed license key T810190 r AJ 46209 46209 en US 160 ...
Страница 179: ...Technical data 27 T199043 FLIR ResearchIR Max 4 Upgrade printed license key T810190 r AJ 46209 46209 en US 167 ...
Страница 213: ...Technical data 27 T199043 FLIR ResearchIR Max 4 Upgrade printed license key T810190 r AJ 46209 46209 en US 201 ...
Страница 220: ...Technical data 27 T199043 FLIR ResearchIR Max 4 Upgrade printed license key T810190 r AJ 46209 46209 en US 208 ...
Страница 248: ...Mechanical drawings 28 See next page T810190 r AJ 46209 46209 en US 236 ...
Страница 250: ...CE Declaration of conformity 29 See next page T810190 r AJ 46209 46209 en US 238 ...
Страница 251: ......
Страница 256: ...Application examples 30 T810190 r AJ 46209 46209 en US 244 ...
Страница 297: ... T810190 r AJ 46209 46209 en US 285 ...
Страница 299: ......