FLIR
CM82
USER
MANUAL
Document
Identifier:
CM82
‐
en
‐
US_AD
19
5.8.2
Three
‐
Phase
Power
Measurements
5.8.2.1
Three
‐
Phase
Three
‐
Wire
Balanced/Unbalanced
The
power
of
a
three
‐
phase
three
‐
wire
wye
(star)
configuration
is
measured
in
two
steps,
in
accordance
with
Figure
5.4.
The
total
power
is
the
sum
of
the
two
measurements:
W
=
W
1
+
W
2
.
Figure
5.4
Three
‐
phase
three
‐
wire
measurements
1.
Set
the
function
switch
to
the
W
position.
2.
Ensure
that
the
meter
is
set
to
active
power
measurement.
If
the
or
the
indicator
is
displayed,
press
the
MODE
button
repeatedly
until
none
of
these
indicators
are
displayed.
3.
Take
two
measurements
of
the
active
power,
in
accordance
with
Figure
5.4.
4.
To
measure
and
display
the
power
factor,
press
the
MODE
button
repeatedly
until
the
indicator
is
displayed.
5.
Read
the
power
factor
value
on
the
display.
If
the
value
is
displayed
without
a
sign,
the
phase
of
the
current
signal
is
lagging
behind
the
voltage
signal
(inductive
load).
If
the
value
is
displayed
with
a
minus
sign
(–),
the
phase
of
the
current
signal
is
leading
the
voltage
signal
(capacitive
load).
6.
To
return
to
active
power
measurements,
press
the
MODE
button
repeatedly
until
neither
the
nor
the
indicator
is
displayed.