5 –Camera Controller
A6600/A6650 User’s Manual
24
Correction (NUC) Types
One Point
Sets the gain terms to “1” and computes the offset terms.
Uses a single NUC source. Does not compute a BP
correction.
Two Point
Sets both the gain and offset terms. Uses two NUC
sources. Computes a bad pixel correction.
Offset Update
Retains the current NUC gain terms and updates the offset
terms. Uses a single NUC source. Retains the current bad
pixel (BP) correction.
Correction Sources
Internal Flag
Use the internal flag as the NUC source. Because the flag
is not temperature controlled, it can only be used for 1-point
and Offset Update NUC functions
External Blackbody
Use an external blackbody as the NUC source. Program
will prompt the user to place each source in front of the
camera. NUC source needs to fill the entire field of view.
Number of frames
Set the number of to average when computing NUC
coefficients. 16 frames is the default and works well for
most scenarios. The value can be to be 2, 4, 8, 16, 32, 64,
or 128.
After configuring the correction parameters and selecting
Next>>
the next window allows the user to
set up the parameters used for the Bad Pixel Detection. Once the parameters are set, select
Next>>