Kabelanschlusskasten Typ KPK-4/L6 Technisches Handbuch, Rev. 1.03 April 2017
Junction Box Type KPK-4/L6 Technical Manual
Seite / Page 11/12
Load Cell Cable Connection
First the cable gland (see (1) in fig.1) must be loosened. Then you have to feed the load cell cable through the
cable gland unless the shrink tube is fully disappeared in the box. The wires have to run below the printed
circuit board and will be pulled back to the top at the upper end of the printed circuit board. Afterwards you can
connect the cables to the clamping terminals as indicated below:
Cable colour
Description
Terminal designation
Flintec colour code
yellow
Cable shield
Schirm / Shield
brown
Sense -
Sense - / Rückführung -
black
Excitation - / ( Input - )
Excitation - / Speisung -
blue
Sense +
Sense + / Rückführung +
green
Exci ( Input + )
Exci / Sp
red
Signal
– ( Output - )
Signal -
white
( )
After all conductors have been clamped to the terminals, the cable glands must be tightened. Please verify
that all cable glands are tight and the cable is fully stress relieved.
Output Cable Connection
The signal cable (connection between junction box and the following electronics) should be a 6
– wire
shielded cable and must be kept as short as possible. Because this cable usually has a larger diameter than
the load cell cable, you should use the larger cable gland in the middle of the box. Depending on type and
manufacturer signal cables may have different colours. Therefore, make your own choice.
Cable colour, example
Description / Terminal designation
outer cable screen
= Shield
pink
= Signal
– (Output –)
white
= ()
grey
= Sense
–
brown
= Excitation
– (Input –)
yellow
= Sense +
green
= Exci (Input +)
Figure 2.
PCB KPK-4 / L6 Terminal designation
0 Ohm jumpers 4x
In case of corner correction will be replaced with resistors
Surge arrestors