Operator's Manual
|
6-16
Shake off excess water and place all parts on a clean
towel to air dry. (Do not heat or blow dry.)
To disinfect the Humidifier Temperature Probe:
1. Soak the probe tips in one part white vinegar to one part of water for 2
hours.
2. Rinse thoroughly with sterile, distilled water.
3. Air dry.
Alternatively:
1. Place the probe tips in a glutaraldehyde solution such as Cidex (2%) for 2
hours.
2. Rinse thoroughly with sterile, distilled water.
3. Air dry.
To sterilize the Humidifier Temperature Probe:
The following method is suitable for sterilizing the entire probe assembly:
The following sterilizing methods are suitable for the bottle and heat sink.
1. Soak in an approved chemical sterilant following the manufacturer’s
instructions.
2. Rinse thoroughly with sterile, distilled water.
3. Air dry.
Ethylene Oxide 55ºC / 131ºF
Visually inspect all parts for cracks or damage. Do not use
cracked or damaged parts. Contact FLIGHT
MEDICAL
for
replacements.
6.1.5 Reusable Patient Circuits
Clean and sterilize patient circuits once weekly while in use. Always use a
clean, disinfected exhalation valve when the patient circuit is reassembled for
patient use.
The breathing circuit includes the tubing, water trap (if used), proximal
pressure line and exhalation valve drive line.
FLIGHT MEDICAL patient circuits are supplied non-sterile.