Flight DEFENDME Aerosol Evacuator R.1.2
17
Recommended minimum separation distances
Nowadays, many RF wireless equipments have being used in various healthcare locations where medical equipment and/or systems are
used. When they are used in close proximity to medical equipment and/or systems, the medical equipment and/or systems’ basic safety
and essential performance may be affected. The
DEFENDME
has been tested with the immunity test level in the below table and meet the
related requirements of IEC 60601-1-2:2014. The customer and/or user should help keep a minimum distance between RF wireless
communications
equipment and the
DEFENDME
as recommended below.
Test
frequency (MHz)
Band
(MHz)
Service
Modulation
Maximum
power (W)
Distance (m)
Immunity
test level (V/m)
385
380-390
TETRA 400
Pulse
modulation 18Hz
1.8
0.3
27
450
430-470
GMRS 460
FRS 460
FM
± 5 kHz deviation 1 kHz
sine
2
0.3
28
710
704-787
LTE
Band 13, 17
Pulse modulation 217Hz
0.2
0.3
9
745
780
810
800-960
GSM 800/900,
TETRA 800,
iDEN 820,
CDMA 850,
LTE Band 5
Pulse modulation 18Hz
2
0.3
28
870
930