Flight DEFENDME Aerosol Evacuator R.1.2
10
display “OFF”. To disconnect the power supply please press the power switch on the main
enclosure box.
7.0
Maintenance
7.1 Cleaning and Disinfection
7.1.1 Cleaning
For hygiene and safety purposes, the suction nozzle must be cleaned and disinfected before
each use to prevent any cross-contamination. The components directional duct, main enclosure,
power cord and speed controller should be cleaned and disinfected regularly according to the
frequency of use. Comply with your local regulations for cleaning and disinfection.
The device should no longer be used if there are signs of damage to the device or any its
Condensation may appear on the suction nozzle if the temperature of the
suction nozzle is much lower than the atmospheric temperature. If this occurs,
run the device for 15-30 minutes continuously without suctioning any aerosols
to make sure there is no water vapour in the directional duct and suction nozzle
before continuing to use.
Components
:
1. Main Enclosure
2. Main Filter Assembly
3. Secondary Filter Assembly
4. Primary Microfiber Filter
5. Rubber Gasket/Seal
6. Top Cover
7. Directional Duct
8. Suction Nozzle