INSTRUCTION HANDBOOK
PD12IHS / PD12IDH
PD12IHS_PD12IDH IH EN 74-115-181 v2.01
Version: 2.01
Page 14 of 19
Function 4 - Acceleration/deceleration
Value: An integral number.
This function offers a choice of values between 1 and 200 dependent on the tube size and drive;
1 = slowest, 200 = fastest.
Tube Sizes
Max. Velocity
Max. acceleration
0.5 - 0.8
– 1.2 – 1.6
400
100
3.2
400
100
4.8 - 6.0
– 8.0
400
100
Function 5 - Reversing (back suction)
Value: An integral number.
After each filling the dispenser head can be set to perform a small back suction to prevent dripping.
The back suction can be set at values between 0 and 10.
0 = no back suction
10 = maximum back suction
The value has no relation to any other parameters and is solely a number of degrees of a rotor turn.
Consequently, the volume that is sucked back will depend on the tube diameter.
For other program possibilities; see the MC12 manual.