Inside   the   amplifier   case   exist   dangerous   voltage   levels.

The   instrument   cannot   be   powered   from   a   DC-­‐AC   converter   nor   

from   a   solid-­‐state   AC   generator   with   non-­‐sinusoidal   output.

Loads   sensi@ve   to   voltage   transients   should   be   disconnected   

from   the   amplifier   during   power-­‐up   and   power-­‐down.

Never   connect   the   output   to   the   input   of   the   amplifier!

The   amplifier   may   be   overheated   if   the   output   is   short-­‐circuited   

for   a   long   @me.

The   maximum   allowable   capaci@ve   load   depend   on   the   internal   

seSng   of   the   slew   rate.   Overloading   the   output   is   likely   to   cause   

overshoot.   Slow   down   the   amplifier   to   accommodate   a   larger   

load.

It   is   recommended   to   monitor   the   output   signal   of   the   amplifier   

on   the   oscilloscope.

ELECTRONICS AB

WWW.FLCE.SE

Содержание A400X

Страница 1: ...BAND 200V 20x DC to ca 500 kHz 150mA FULL SCALE LOW OUTPUT IMPEDANCE HIGH SLEW RATE 0 1 400 V s ELECTRONICS AB WWW FLCE SE FLC Electronics AB Sippedalsv gen 8 S 433 31 Partille Sweden Tel 46 31 741 26...

Страница 2: ...used separately allow the unit to act as inverting or non inverting amplifier Any function or arbitrary waveform generator or any other signal source with a voltage up to 10 V can be used as an input...

Страница 3: ...e se for advice before opening the instrument case Keep input signals within 10 V range Never connect the high voltage output to the input of the instrument TROUBLESHOOTING Problem Condi on Solu on No...

Страница 4: ...preferably by email info flce se Inside the cabinet exist hazardous voltage levels and the amplifier circuit is sensitive to static discharge FLC Electronics AB recommends to monitor the output signal...

Страница 5: ...cy response without load and with 300pF and 400 pF load respectively is shown in the following diagram 1 10 100 1000 1 10 100 1000 10000 Amplitude peak to peak V Frequency kHz without load load 300pF...

Страница 6: ...max 150 mA Operating Ambient Temperature 0 C to 30 C Storage Temperature 0 C to 60 C Relative Humidity up to 90 operation 30 to 50 storage Power Requirements 100 110 V or 220 230 V 50 60 Hz Fuse 100 1...

Страница 7: ...furnish service under this warranty to repair damage resulting from attempts by personnel other than FLC Electronics representatives to install repair or service the product to repair damage resulting...

Страница 8: ...li er during power up and power down Never connect the output to the input of the ampli er The ampli er may be overheated if the output is short circuited for a long me The maximum allowable capaci ve...

Страница 9: ...d to the following standards as listed in the official Journal of the European Communities EN 50081 1 Generic Emissions EN 55022 Conducted emission interference voltage class B EN 55022 Radiated emiss...

Отзывы: