TROUBLESHOOTING
Switch off the device before any procedure and unplug the power cable from the socket.
PROBLEM
CAUSE
SOLUTION
The device does not work
The power cable has not been cor-
rectly inserted into the power socket
Correctly insert the power cable in sockets.
The device does not
nebulize or
nebulizes insufficiently.
The medication has not been inserted
in the nebulizer
Pour the right amount of medication in the
nebulizer
The nebulizer has not been properly
fitted.
Disassemble the nebulizer and
reassemble it correctly as shown in the connection
diagram on the cover.
The nebulizer nozzle is clogged
Disassemble the nebulizer, remove the nozzle and
perform cleaning operations.
Failure to remove medication deposits from the
nebulizer affects its efficiency and operation.
Strictly comply with the instructions contained
in the “CLEANING, SANITISATION DISINFECTION”
chapter
The air pipe is not correctly connected
to the device
Ensure the device’s accessories
are properly connected to the air intake (see
connection diagram on the cover).
The air pipe is bent, damaged or
twisted
Unwind the pipe and ensure it is
not crushed or punctured. If necessary replace it.
The air filter is dirty
Replace the filter
The apparatus is noisier
than usual
The Filter is not correctly
inserted in its housing
Fully and properly insert the Filter in its housing
If after verifying the above mentioned conditions the device should not operate properly, contact your trusted retailer or
the nearest authorized FLAEM service center.
SYMBOLS
CE Medical Marking ref. Dir. 93/42 EEC and
subsequent amendments
Serial number of device
Class II device
Manufacturer
Important: check the operating instructions
Type BF applied part
“OFF” for part of equipment
Alternating current
“ON” for part of equipment
Complies with: European standard EN 10993-
1 “Biological Evaluation of Medical Devices”
and European Directive 93/42/EEC “Medical
Devices”. Phthalates free. In conformity with
Reg. (EC) no. 1907/2006
IP21
Enclosure protection rating: IP21.
(Protected against solid bodies over 12 mm.
Protected against access with a finger; Protected
against vertically falling water drops).
Minimum and maximum room temperature
Minimum and maximum air moisture
Minimum and maximum atmospheric pressure